4. Review the A rrhythmia Logbook screen and for episodes of interest, print episode details and stored
electrogram information.
5. Clear the counters and histograms so that the most recent episode data will be displayed at the next
follow-up session.
NOTE: Echo-Doppler studies may be used to non-invasively evaluate AV Delay and other programming
options post-implant.
CAUTION: Verify with a conversion test that the patient’s tachyarrhythmias can be detected and terminated
by the pulse generator system if the patient’s status has changed or parameters have been reprogrammed.
EXPLANTATION
NOTE: Return all explanted pulse generators and leads to Boston Scientic. Examination of explanted pulse
generators and leads can provide information for continued improvement in system reliability and warranty
considerations.
WARNING: Do not reuse, reprocess, or resterilize. Reuse, reprocess ing, or resterilization may compromise
the structural integrity of the device and/or lead to device failure which, in turn, may result in patient injury,
illness, or death. Reuse, reprocessing, or resterilization may also create a risk of contamination of the device
and/or cause patient infection or cross-infection, including, but not limited to, the transmission of infectious
disease(s) from one patient to another. Contamination of the device may lead to injury, illness, or death of
the patient.
Contact Boston Scientic when any of the following occur:
When a product is removed from service.
In the event of patient death (regardless of cause), along with an autopsy report, if performed.
For other observation or complications reasons.