TACHYARRHYTHMIA DETECTION
VENTRICULAR DETECTION
3-39
Table 3-14. Combin atio ns o f Onset And Stability and resulting therapy (continued)
Detection
Rhythm
Onset An d Stability C om bination
ab
Onset Or Stability Combination
c
Sudden, unstable
Inhibit Treat
Sudden, stable
Treat Treat
a. If the detected ventricular rhythm changes, then the appropriate, corresponding row in the table is evaluated.
b. The And combination is the nominal setting when both are enabled.
c. Decisions to inhibit can be overridden by V > A or expiration of SRD.