User Manual Part 1
USING THE PROGRAMMER/RECORDER/MONIT
OR
INDICATIONS-BASED PROGRAMMING (IBP)
2-3
Figure 2-1. Indications-base d Programming screen
Indications are clustered in general categories as illustrated abo ve. The inten t
for each category of indications is described below:
• Sinus Node:
– If Normal is selected, the intent is to provide RV pacing when necessary.
– If C hronotropically Incompetent is selected, the intent is to provide
rate-adaptive pacing.
– If Sick Sin us Syndrome is selecte d, the intent is to provide atrial pacing
support.
• AV Node:
– If Normal or 1st Degree Block is selected, the intent is to provide RV
pacing when necessary.
– If 2nd Degree Block is selected, the intent is to promote intrinsic AV
conduction and provide AV sequential pacing when conduction is not
present.
– If Complete Heart Block is selected, the intent is to provide AV
sequential pacing.
NOTE: The selected settings for AF and Sinus Node may affect the suggested
value for the Normal/1st Degree Block setting of AV Node.