Conduct diagnostic tests including the following:
Pacing im pedance tests
Pacing threshold tests
Intrinsic amplitude tests
Perform manual capacitor re-form
ZIP Telemetry
ZIP telemetry is a wandless, two-way RF communication option that allows the
PRM system to communicate with some pulse generator models. W hen a
wanded telemetry session is initiated, the PRM checks the pulse generator’s
telemetry capability. If the PRM detects a pulse ge n era tor with ZIP telemetry
capability, a message will display indicating that ZIP telemetry is a va ilab le and
the wand can be removed. Otherwise, the session will continue with wanded
telemetry.
ZIP telemetry offers the following advantages over traditional wanded telemetry:
The faster data transmission speed means less time is required for device
interrogation
Data transmission over a longer distance (within 3 meters) minimizes the
need to keep the wand in the sterile eld during implant, which may reduce
the risk of infection
Continuous telemetry is possible during the entire implant procedure,
allowing monitoring of pulse generator performance and lead integrity
during implant
Regardless of whether ZIP telemetry is being used, wanded communication
is still available.
Starting a Wanded Telemetry Session
Follow this procedure to begin a wanded telemetry communication session:
1. Make sure the telemetry wand is connected to the PRM system and is
available throughout the session.
2. Position the wand over the pulse generator at a distance not greater than 6
cm (2.4 inches).