User Manual Part 1

3-20
TACHYARRHYTHMIA DETECTION
VENTRICULAR DETECTION
Table 3-6. End-of-Episode Timer (continued)
Episode Classication Ventricular En d- of -Episode Timer
(elapsed time requ ired to d eclare episode over)
Treated (only ATP therapy delivered)
10 seconds
Treated (any shock therapy delivered)
30 seconds
NOTE: The episode is terminated immediately if the Tachy Mode is
reprogrammed, an inductio n m e thod or lead test is attempted b ef ore the
End-of-Episode time out, or any ventricular detection or ventricular therapy
parameters are reprogrammed.
Coupling
Interval
Detection window does
not become satisfied.
End-of-Episode times out.
Episode is over.
Detection met.
Deliver ATP.
Start Duration.
Start episode.
Start End-of-
Episode timer.
Detection window remains satisfied.
ATP Burst
10 s
Detection window satisfied.
Figure 3-12. Treated episode, ventricular mode is Monitor + Th erapy and ATP is delive red
Detection window remains satisfied.
Detection met.
Start Charging.
Start Duration.
Start episode.
Detection window satisfied.
Charging
Shock
End-of-Episode times out.
Episode is over.
Start End-of-Episode timer.
30 s
Detection window does not become satisfied.
Figure 3-13. Treated episode, ventricular mode is Monitor + The rapy and shock is delivered
- DRAFT -