User Manual Part 2

TAC HYARRHYTHMIA THERAPY
VENTRICULAR SHOCK THERAPY AND PARAM ETER S
4-21
VENTRICULAR SHOCK THERAPY AND PARAMETERS
The pulse generator delivers shocks synchronous to a sensed event. The
shock vector, energy level, and polarity of the shocks are programmable.
Ventricular Shock Vector
The programmed Ventricular Shock Vector indicates the vector of energy
delivery for ventricular shock therapy.
The following programmable congurations are available:
RV Coil to RA Coil and Can––this vector is also known as the V-TRIAD
vector. It uses the metallic hou sing of the pulse generato r as an ac tive
electrode (“hot can”) combined w ith the ENDOTAK two-electrode
debrillation lead. Energy is sent via a dual-current pathwa y fro m the dista l
shocking electrode to the proximal electrode and to the pulse generator
case.
RV C oil to Can––this vec to r uses the metallic housing of the pulse
generator as an active electrode (“hot can”). Energy is sent from the distal
shocking electrode to the pulse generator case. This conguration should
be selected when using a single-coil lead.
RV Coil to RA Coil––this vector removes the pulse generator case as
an active electrode and is also known as a “cold can” vector. Energy is
sent from the distal shocking ele ctrode to the proximal ele ctrode. This
vector shou ld never be used with a single-coil lead, as a sh o c k will not
be delivered.
Ventricular Shock Energy
Ventricular shock energy determines the strength of shock therapy delivered
by the pulse generator.
Shock output remains constant over the lifetime of the pulse generator,
regardless of changes in lead impedance o r battery voltage. The constant
output is accomplished b y varying pulse width to adjust to changes in lead
impedance.
The rst two shocks in each ventricular zone can be programmed to optimize
charge time, longevity, and safety margins. The remaining shock energies in
each zone are nonprogrammable at the maximum-energy value.
- DRAFT -