User Manual Part 2

TAC HYARRHYTHMIA THERAPY
VENTRICULAR SHOCK THERAPY AND PARAM ETER S
4-23
CAUTION: For IS-1/DF-1 leads, never change the shock waveform
polarity by physically switching the lead anodes and cathodes in the pulse
generator header—use the programmable Polarity feature. Device damage
or nonconversion of the arrhythmia post-operatively may result if the polarity
is switched physically.
Biphasic
V1
V2
PW1
PW2
V3
V4
PW = Pulse Width
PW2 = PW1 x 0.66
V2 = V3
Figure 4-21. Biphasic waveform
+
+
+
Initial polarity Reverse polarity
Figure 4-22. Polarity of shock deliv ery
Committed Shock/Reconrmation of th e Ventricular Arrhythmia
Committed Shock/Reconrm ation refers to the monitoring performed by the
pulse generator before delivery of a ventricular shock.
If the patient is subject to non-sustained a rrhythmias, rec onrmation may be
desirable in order to prevent delivery of unnecessary shocks to the patient.