PACING THERAPIES
BASIC PARAMETERS
5-7
STAT PACE––initiates emergency ventricular pacing at high output settings
when commanded via the PRM using telemetry communication.
BASIC PARAMETERS
By programming device parameters, the pulse generator provides CRT for the
intent of providing mechanical synchronization. The programming options used
for CRT include those used for bradycardia pacing therapy.
LV stimulation is delivered using a Guidant coronary venous lead. The device
uses bipolar atrial pacing and sensing to coordinate AV contractions with CRT.
Normal Settings include the following:
Pacing parameters, which are independently programmable from
post-therapy and temporary pacing parameters
NOTE: These parameters are also used for CRT.
Pacing and Sensing
Leads
Sensors and Tren din g
Post-Therapy Settings include the following:
Pacing parameters, which are independently programmable from normal
and temporary pacing parameters
Post-ventricular sh ock
Brady Mode
Brady modes provide you with programmable options to help individualize
patient therapy.
This pulse generator includes the pacing modes identied in the Programmable
Options appendix.
CRT Modes
The objective of CRT is to deliver continuous pacing to the ventricles. CRT can
only be delivered in modes that provide ventricular pacing.