User Manual Part 2

PACING THERAPIES
REFRACTORY
5-57
If the value is programmed to Smart, the pulse generator automatically
adjusts the sensit ivity va lue in order to reject far-eld atrial events. This
allows for sensing of true ventricular events that had previously fallen in the
cross-chamber blanking period.
LV-Blank after A-Pace
LV-Blank after A-Pace, a cross chamber blanking period, inhibits LV sensing
following an atrial pace.
If the value is programmed to Smart, the pulse generator automatically
adjusts the sensit ivity va lue in order to reject far-eld atrial events. This
allows for sensing of true ventricular events that had previously fallen in the
cross-chamber blanking period.
A-Blank after V-Pace
A-Blank after V-Pace, a cross-chamber blanking period, inhibits atrial sensing
following an RV or LV ventricular pace.
If the value is programmed to Smart, the pulse generator automatically adjusts
the sensitivity value in order to reject far-eld ventricular events. This allows
for sensing of true atrial events that had previously fallen in the cross-chamber
blanking period .
A-Blank after RV-Sense
A-Blank after RV-Sense, a cross-chamber blanking period, inhibits atrial
sensing following an RV sensed event.
If the value is programmed to Smart, the pulse generator automatically adjusts
the sensitivity value in order to reject far-eld ventricular events. This allows
for sensing of true atrial events that had previously fallen in the cross-chamber
blanking period .
Refer to the following illustrations:
- DRAFT -