User Manual Part 1
INFORMATION FOR USE
PRECAUTIONS
1-9
•STATPACEsettings.When a pulse generator is programmed to STAT
PACE settings, it will continue to pace a t the high-energy STAT PACE
values if it is not reprogrammed. The use of STAT PA CE parameters will
decrease dev ice longevity.
• Biventricular pacing therapy. This device is intended to provide
biventricular pacing therapy. Programming the device to provide RV-only
pacing, or programming the RV pace amplitude below the pacing threshold
(resulting in LV-only pacing), is not intended for the treatment of heart
failure. The clinical effects of LV-only or RV-only pacing for the treatment of
heart failure have not been established.
• Pacing and sensing margins. Consider lead maturation in your choice of
pacing amplitude, pacing pulse width, and sensitivity settings.
• An acute pacing threshold greater than 1.5 V or a chronic pacing
threshold greater than 3 V can result in loss of capture because
thresholds may increase over time.
• An R-wave amplitude less than 5 mV or a P-wave amplitude less than
2 mV can result in undersensing because the sensed amplitude may
decrease after implan ta tio n.
• Pacing lead impedance should be within the range of 200 and
2000 .
• Line-powered equipment. Exercise extreme caution if testing leads u sing
line-powered equipment because leakage current exceeding 10 µA can
induce ventricular fibrillation. Ensure that any line -p owered equipment is
within specifications.
• Proper programming of the lead configuration. If the Lead Configu rat ion
is programmed to Bipolar when a unipolar lead is implanted, pacing will
not occ ur.
• Proper programming of the shock vector. If the shoc k vector is
programmed to RVcoil>>RAcoil and the lead does not have an RA coil,
shocking will not occur.
• Replacement device. Implanting a replacement device in a subcutaneous
pocket that previously housed a larger device may result in pocket air
entrapment , migration, erosion, or insu fficient grounding between the
device and tissue. Irrigating the pocket with sterile saline solution decreases
the pos sibility of pocket air entrapmen t and in su fficient grounding. Suturing
the de vice in place reduces the possibility of migration and erosion.
- DRAFT -