INFORMATION FOR USE
X-RAY IDENTIFIER
Table 1-3. Characteristics as shipped (continued)
Parameter
Setting
Pacing Therapy available DDDR
Sensor Accelerometer (MV for respiratory rate trend)
Pace/Sense Conguration RA: BI/BI
Pace/Sense Conguration RV: BI/BI
Pace/Sense Conguration LV: Off
The pulse generator is shipped in a power-saving Storage mode to extend its
shelf life. In Storage mode, all features are inactive except:
Telemetry support, which allows interrogation and programming
Real-time clock
Commanded capacitor re-formation
STAT SHOCK and STAT PACE commands
The device leaves Storage mode when one of the following actions occurs;
however, programming other parameters will not affect the Storage mode:
STAT SHOCK or STAT PACE is commanded
Tachy Mode is programmed to:
–Off
Monitor Only
Monitor + Therapy
Once you have programmed the pulse generator out of Storage mode, the
device cannot be reprogrammed to that mode.
X-RAY IDENTIFIER
The pulse generator has an identier that is visible on x-ray lm or under
uoroscopy. This identier provides noninvasive conrmation of the
manufacturer and consists of the following:
The letters, BOS, to identify Boston Scienticasthemanufacturer
The number, 112, to identify the Model 2868 PRM software application
needed to communicate with the pulse generator