Charging frequency
Longevity is also reduced in the following circumstances:
With a decrease in pacing impedance
When Patient Triggered Monitor is programmed to On
For models with ZIP wandless telemetry, one hour of additional telemetry
reduces longevity by approximately 4 days.
Device longevity may also be affected by:
Tolerances of electronic components
Variations in programmed paramete r s
Variations in usage as a result of patient condition
An additional maximum-energy shock reduces longevity by approximately
19 days.
Refer to the PRM Summary screen for an estimate of pulse generator longevity
specic to the implanted device.
WA RRANT Y INFORMATION
A limited warranty certicate for the pulse generator is packaged with the
device. For additional copies, please contact Boston Scientic at the address
and phone number shown on the back cover of this manual.
PRODUCT RELIABILITY
It is Boston Scientic’s intent to provide implantable devices of high quality and
reliability. However, thes e devices may exhib it malfu n ctio ns that may result in
lost o r compromised ab ility to deliver ther ap y. These malfunctions may include
the following:
Premature battery depletion
Sensing or pacing issues
Inability to shock
Error codes
Loss of telemetry