INDICATIONS-BASED PROGRAM MING (IBP)
2-3
Figure 2-1. Indications-based Programming screen
Indications are clustered in ge ne r al categ ories as illustrated above. The intent
for each category of indications is described below:
Sinus Node:
If Normal is selected, the intent is to promote intrinsic atrial events
and provide CRT pacing.
If Chronotropically Incompetent is selected, the intent is to provide
rate-adaptive CRT pacing.
If Sick Sinus Syndrome is selected, the intent is to prov id e atrial pa c in g
support and CRT pacing.
AV Node:
The intent is for nominal Paced AV delay and Sensed AV delay
settings. The SmartDelay optimization feature may be used to adjust
the AV delay.
NOTE: The selected settings for AF and Sinus Node may affect the suggested
value for the setting of AV Node.
Atrial Arrhythmias
If Paroxysmal/P ersistent is selected, the intent is to avoid tracking
atrial arrhythmias by using dual-chamber pacing mode with ATR Mode
Switch.
If Permanent/Chronic AF is selected, the intent is to provide rate
adaptive RV pacing.