User Manual Part 1

TACHYARRHYTHMIA DETECTION
VENTRICULAR DETECTION
3-9
If the passive method is enabled, the pulse generator will attempt to collect the
Rhythm ID reference template every two hours using the programmed brady
settings .
Iftheactivemethodisenabledandsevendayshavepassedsincethelast
successful collection of a reference template, then every 28 hours the device
automatically analyzes the patient’s intrinsic rhythm by adjusting the brady
parameters. During a Rhythm ID active reference template update, the
following will occur:
1. The device veries that the patient is at rest (as measured by the
accelerometer input).
2. The device enables a controlled pacing rate decrease to the programmed
Rhythm ID Fallback LRL. Du r ing this f allb ack period, the follo win g occurs:
The device temporarily switches the pacing m ode to DDI, VDI, VVI,
AAI, or Off (according to the programmed brady mode) and extends
the AV delay up to 400 ms.
Biventric ular Trigger, Rate Smoothing, ATR, hysteresis, and dynamic
programming (excluding Dynamic VRP) are suspended. The pacing
chamber is set to Biventricular; LV Offset is set to 0.
3. After the Fallback period, pacing parameters are restored to normal
programmed parameters. Fallback periods occur no m ore than once per
day and will typically last less than one minute.
A method for manually commanding the device to acquire a Rhythm ID
referencetemplateisalsoavailable.
NOTE: If Rhythm ID is not enabled, a manual reference template update can
still be performed. However, the acquired template will not be used in analysis
to determine if the patient’s rhythm is VT or SVT.
During a manual Rhythm ID reference template update, the pulse generator will
perform the following tasks:
1. Enable a controlled rate decrease to t he programmed Rhythm ID Fallback
LRL. During the fallback period, the following occurs:
The device temporarily switches to the programmed Manual Rhythm
ID Brady Mode and extends the AV delay up to 400 ms.
- DRAFT -