User Manual Part 2

PROGRAMMABLE OPTIONS
A-11
Table A-12. Pacing therapy parameters (Normal, Post-T herap y, and Temporary) (specie d into a 750
load) (cont inued)
Parameter Programmable Values Nom inal
Maximum Sensed AV D elay
ag
(ms)
30, 40, ..., 400
150(Tolerance±5ms)
Minimum Sensed AV Delay
ag
(ms)
30, 40, ..., 400
65 (Tolerance ± 5 m s)
AV Search +
gj
Off, On Off
AV Search + Search Interval
gj
(cycles)
32, 64, 128, 256, 512, 1024
32 (Tolerance ± 1 cycle)
AV Search + Search AV Delay
gj
(ms)
30,40 ..., 400
300(Tolerance±5ms)
Respiratory Sensor
ag
Off, On On
Rate Hysteresis Hysteresis O ffset
g
j
( ppm)
-80, -75, ... ,-5, Off Off (Tolerance ± 5 ms)
Rate Hysteresis Search Hysteresis
g
j
(cycles)
Off, 256, 512, 1024, 2048, 4096 Off (Tolerance ± 1 cycle)
Rate Smoothing (up, down)
gj
(%) Off, 3, 6, 9, 12, 15, 18, 21, 25 Off (Tolerance 1%)
Noise Response
agl
AOO, VOO, DOO , Inhibit Pacing DOO for DDD(R) and DDI(R)
modes; VOO for VDD(R) and VVI(R)
modes; AO O for AAI(R) mod e
Maximum Pacing Rate
ag
(ppm)
30, 35, ... ,185
130(Tolerance±5ms)
Post-therapy Pacing Period
(min:sec) (available post-shock
only)
00:15, 00:30, 00:45, 01:00, 01:30,
02:00, 03:00, 04:00, 05:00, 10:00,
15:00, 30:00, 45: 00, and 60:00
00:30 (Tolerance ± 1 cardiac cycle)
a. The programmed Normal Brady values will be used as the nominal values for Temporary B rady pacing.
b. Refer to the NASPE/BPEG codes below for an explanation of the programmable values. The identication code of the North
American Society of Pacing and Electrophysiology (NASPE) and the British Pacing and Electrophysiology Group (BPEG) is based
on the categories listed in the table.
c. The basic pulse period is equal to the pacing rate and the pulse interval (no hysteresis). Runaway protection circuitry inhibits
bradycardia pacing above 205 ppm. Magnet application does not affect pacing rate (test pulse interval).
d. Separately programmable for ATP/Post-Shock, Temporary Brady, and EP Test.
e. The minimum value of energy delivered at 5 V and 0.5 ms is 20 µJ with 200–500 , and 12 µJ with 1000 resistive load
at 37°C ± 1°C for BOL and Explant.
f. Values are not affected by temperature variation within the range 20°–43°C.
g. This parameter is used globally in Normal Brady pacing and Post-shock Brady pacing. Changing the value for Normal Brady will
change the value for Post-shock Brady.
h. This parameter is automatically set to at least 85 ms for Post-Shock Brady.
i. This parameter is automatically adjusted in Post-Shock Brady to allow appropriate sensing.
j. This parameter is disabled during Temporary Brady.
k. The programmable values for VR devices only include VVI(R), Off; Temporary: VVI, VOO, Off
l. The programmable values for VR devices only includes VOO and Inhibit Pacing and as such the nominal is VOO.
Table A-13. Atrial Tachy Parameters
Parameter Programmable Values Nom inal
ATR Mode Switch
ab
Off, On On
ATR Trigger Rate
ab
( bpm)
100, 110, ..., 300
170(Tolerance±5ms)
- DRAFT -