User Manual Part 2

PROGRAMMABLE OPTIONS
A-13
Table A-15. Ma gn et and B eeper functio ns
Parameter Programmable Values Nom inal
Magnet Response
Off, Store EGM, In hibit Therapy
Inhibit Therapy
Beep During C apacitor Ch arge Off, On Off
Beep When Explant is Indicated
Off, On On
Table A-16. Sensitivity Adjustment
Parameter Programmable Values Nom inal
Atrial Sensitivity AGC 0.15, AGC 0.2, AGC 0.25,
AGC 0.3, AGC 0.4, ..., AGC 1.0,
AGC 1.5
AGC 0.25
Right Ventricular Sensitivity AGC 0.15, AGC 0.2, AGC 0.25,
AGC 0.3, AGC 0.4, ..., AGC 1.0,
AGC 1.5
AGC 0.6
Table A-17. Ventricular Com m and ed ATP
Parameter
a
Programmable Values Nominal
Commanded Ventricular ATP (Type) Burst, Ramp, Scan, Ramp/Scan
Burst
Number Of Bursts
1, 2, ..., 30 30
Initial Pulses per Burst (pulses)
1, 2, ..., 30 4
Pulse Increment (pulses)
0, 1, .., 5 0
Maximum Number of Pu lses
1, 2, ..., 30 4
Coupling Interval (% or ms)
50, 53 , 56, 59; 63, 66, ..., 84; 88, 91,
94, 97% or 120, 130, ..., 750 m s
81%(Tolerance±5ms)
Coupling Interval Decrement (m s)
0, 2, ..., 30
0(Tolerance±5ms)
Burst Cycle Length ( BCL) (% or ms)
50, 53 , 56, 59; 63, 66, ..., 84; 88, 91,
94, 97% or 120, 130, ..., 750 m s
81%(Tolerance±5ms)
Ramp Decrement (ms)
0, 2, ..., 30
0(Tolerance±5ms)
Scan Decrement (ms)
0, 2, ..., 30
0(Tolerance±5ms)
Minimum Interval (ms)
120, 130, ..., 400
200(Tolerance±5ms)
a. The ventricular Commanded ATP Pulse Width and Amplitude values are the same as programmed for ventricular ATP therapy.
Table A-18. 50 Hz/Manual Burst Pacing
Parameter
a
Programmable Values Nominal
Burst Interval (ms)
20, 30, ..., 750
600(Tolerance±5ms)