User Manual Part 2

B-1
PACEMAKER INTERACTION
APPENDIX B
There may be instances where patients may have a separate temporary or
permanent pacemaker. Temporary or permanent pacemakers can interact
with an ICD and can interfere with th e ide ntication of tachyarrhythmias in
the following ways:
During a tachyarrhythmia, if the pacemaker does not sense the arrhythmia
and paces, and the pacing pulse detected from the ICD rate-sensing
electrode is large enough, it c ould cause the ICD to interpret the pacing
as a normal rhythm at the rate of the pacemaker. The ICD would neither
detect the arrhythmia nor deliver therapy.
The pacemaker could present signals to the ICD resulting from the
following events:
Inappropriate sensing
Lead dislodgment
Failure to capture
This could cause the ICD’s rate measurement to be faster than the patient’s
actual heart rate. As a result, the ICD could deliver inappropriate therapy.
Conduction delay could cause the ICD to count both pacemaker artifact and
ventricular depolarization. This could result in inappropriate ICD therapy.
For these reasons, the use of a pacemaker that results in pacemaker and ICD
interaction is not recommended. Unipolar pacemakers are contraindicated for
usewithanICD.
Consider the following actions if a separate pacemaker is used:
Always deactivate a patient’s ICD when:
Using temporary bipolar or temporary AV sequential pacing
Reprogramming a separate implanted pacemaker
If cardioversion or debrillatio n is neede d at such times, use an e xternal
debrillato r.
- DRAFT -