User Manual Part 2

6-6
SYSTEM DIAGNOSTICS
LEAD TESTS
Last Delivered Ventricular Shock
When a shock has been delivered to the patient, the following information
from the last shock delivered is stored in the pulse generator’s memory and
displayed on the Battery Detail screen:
•Date
Energy level
•Chargetime
Shocking lead impedance
This does not include auto capacitor re-forms or shocks that may have been
diverted. If a fault condition is encountered (i.e., high or low impedance), the
fault w ill be indicated so that corrective action may be taken.
NOTE: For shocks of 1.0 J or less, the accuracy of the impedance
measurement decreases.
LEAD TESTS
The following lead tests are available (Figure 6-2 on page 6-6):
Pace Impedance
Shock Impedance
Intrinsic Amplitude
Pace T hre sh old
Figure 6-2. Lead Tests screen
Lead tests can be a ccess ed by using the following step s:
- DRAFT -