User Manual Part 2

ELECTROPHYSIOLOGIC TESTING
EP TEST FEATURES
8-3
Figure 8-1. EP Test Screen
The screen provides the following information:
Status messages indicate detection and therapy status and are described
below:
Ventricular episode status—if an episode is occurring, the duration of
the episode is displayed. (If it is greater than 10 minutes, then it is
displayed as > 10:00 m:s).
Ventricular detection sta tu s—if an episo de is occurring , it indicates
whether ventricular detection is in Initial Detection, Redetection, or the
zone in which that detection is met. If no episode is occurring, the
programmer will also display the text Time since last V therapy along
with the continually updated time in minutes (up to 10).
Brady pacing and SRD status.
The type of therapy initiated and the zone.
The status of the therapy such as In progress, Diverted, or Delivered.
Duration timer—Progression of the duration timer is graphically displayed
using a scale. The bar in the scale moves from left to right to show the
percent complete of programmed duration. When duration is expired and
therapy delivery begins, the bar is removed.
Detection status—The status for each programmed detection enhancement
is displayed. When enhancement criteria are met, a mark appears in the
adjacent box.
- DRAFT -