User Manual Part 3

9-6
IMPLANT INFORMATION
IMPLANTING THE PULSE GENERATOR
may help avoid erosion or extrusion in some patients. Verify magnet function
and wanded telemetry to ensure the pulse generator is within acceptable range.
Consider the following situations during the implant the procedure:
If an abdominal implant is suitable, it is recommended that implantation
occur on the left abdominal side.
Tunnel the leads if necessary. If a Guidant tunneler is not used, cap
the lead terminal pins, gently tunnel the leads subcutaneously to the
implantation pocket, and reevaluate the lead s ignals to determine if any of
the leads have been damaged during the tunneling procedure. A Penrose
drain, large chest tube, or tunneling tool may be used to tunnel the leads.
If the lead terminal pins are not connected to a pulse generator at the time
of lead implantatio n, they must be capped before closing the incision.
Step F: Connect the Leads to the Pulse Gen erato r
Lead connections are illustrated below.
+
RV
RA
DF-1
IS-1
BI
DF-1
LV
LV-1
UNI/BI
IS-1
BI
+
RV
RA
DF-1
IS-1
BI
DF-1
IS-1
BI
LV
IS-1
UNI/BI
Figure 9-1. Lead conn ectio ns
Setscrew locations are illustrated below.
- DRAFT -