User Manual Part 3

A-1
PROGRAMMABLE OPTIONS
APPENDIX A
Table A-1. ZIP Telemetry set tings
Parameter Programmable Values Nominal
Communication Mode Enable use of ZIP telemetry (May
require limited use of wand), Use
wand for all telemetry
Enable u se of ZIP telemetry (May
require limited use of wand)
Table A-2. Tachy Mode parameter
Parameter Programmable Values Nominal
Tachy Mode
Off, Monitor Only, Monitor + Therapy,
Enable Electrocaute ry Protection
Storage
Table A-3. Ventricular Zones parameter
Parameter Programmable Values Nominal
Ventricular Z ones 1, 2, 3 2
Table A-4. Detection parameters for 1-zone, 2-zone, and 3-zone congurations
Parameter VT-1 Z one VT Zone VF Z on e Nominal
Rate
a
( bpm) 3 zones
(intervals in ms)
90, 95, ..., 200
(667–300)
110, 115, ..., 210
(545–286) 22 0 (273)
130, 135, ..., 210
(462–286), 220, 230,
240, 25 0 (273–240)
140 (Tolerance ± 5
ms) for VT-1 Zone
160 (Tolerance ± 5
ms) for VT Zone
200 (Tolerance ± 5
ms) for VF Zone
Rate
a
( bpm) 2 zones
(intervals in ms)
––
90, 95, ..., 210
(667–286) 22 0 (273)
110, 115, ..., 210
(545–286) 220, 230,
240, 25 0 (273–240)
160 (Tolerance ± 5
ms) for VT Zone
200 (Tolerance ± 5
ms) for VF Zone
Rate
a
( bpm) 1 zone
(intervals in ms)
–– ––
90, 95, ..., 210 (667–
286) 220 (273)
200 (Tolerance ± 5
ms)
Initial Duration
b
(sec) 3 zones
1.0, 1.5, ..., 5.0, 6.0,
7.0, ..., 15.0, 20.0,
25.0, ..., 60.0
1.0, 1.5, ..., 5.0, 6.0,
7.0, ..., 15.0, 20.0,
25.0, 30.0
1.0, 1.5, ..., 5.0, 6.0,
7.0, ..., 15.0
2.5 (Tolerance ± 1
cardiac cycle) for
VT-1 Zone
2.5 (Tolerance ± 1
cardiac cycle) for VT
Zone
1.0 (Tolerance ± 1
cardiac cycle) for VF
Zone