User Manual Part 3

A-4
PROGRAMMABLE OPTIONS
Table A-6. Onset/Stability d etection enhancement pa ra meters for 2-z one and 3-zone congurations
(continued)
Parameter VT-1 Z one VT Zone VF Z on e Nominal
Detection
Enhancement 2
zones
––
Off, On
––
On
Atrial
Tachyarrhythmia
Discrimination 3
zones
Off, On
–– ––
On
Atrial
Tachyarrhythmia
Discrimination 2
zones
––
Off, On
––
On
Sinus Tachycardia
Discrimination 3
zones
Off, On
–– ––
On
Sinus Tachycardia
Discrimination 2
zones
––
Off, On
––
On
Polymorphic VT
Discrimination 3
zones
––
Off, On
––
On
Polymorphic VT
Discrimination 2
zones
––
Off, On
––
Off
Table A-7. Rhythm ID detection enhancement parameters for 2-zone and 3-zone congurations
Parameter VT-1 Z one VT Zone VF Z on e Nominal
Detection
Enhancement 3
zones
Off, On Off, On
––
On (VT-1) Off (VT)
Detection
Enhancement 2
zones
––
Off, On
––
Off
Sustained Rate
Duration (min:sec) 3
zones
Off, 00:10, 00:15,
..., 01:00, 01:15,
..., 02:00, 02:30,
..., 10:00, 15:00, ...,
60:00
Off, 00:10, 00:15,
..., 01:00, 01:15,
..., 02:00, 02:30,
..., 10:00, 15:00, ...,
60:00
––
03:00 (VT-1 and
VT) (Tolerance ± 1
cardiac cycle)
Sustained Rate
Duration (min:sec) 2
zones
––
Off, 00:10, 00:15,
..., 01:00, 01:15,
..., 02:00, 02:30,
..., 10:00, 15:00, ...,
60:00
––
03:00 (Tolerance ± 1
cardiac cycle)