PROGRAMMABLE OPTIONS
A-13
Table A-13. B rady/CR T left ventricular pacing parameters (specied into a 750 load) (continued)
Parameter Programmable Values Nominal
a
Left Ventricular Pace Conguration
b
Single or Dual:
LVtip>>Can
LVtip>>RV
Dual Only:
LVring>>Can
LVring>>RV
LVtip>>LVring
LVring>>LVtip
Single: LVtip>>RV
Dual: LVtip>>LVring
Left Ventricular Sense
Conguration
b
Single or Dual:
LVtip>>Can
LVtip>>RV
Off
Dual Only:
LVring>>Can
LVring>>RV
LVtip>>LVring
Single: LVtip>>RV
Dual: LVtip>>LVring
a. The programmed Normal Brady values will be used as the nominal values for Temporary Brady pacing.
b. This parameter is used globally in Normal Brady pacing and Post-shock Brady pacing. Changing the value for Normal Brady will
change the value for Post-shock Brady.
c. Separately programmable for ATP/Post-Shock, Temporary Brady, and EP Test.
d. The minimum value of energy delivered at 5 V and 0.5 ms is 20 µJ with 200–500 , and 12 µJ with 1000 resistive load
at 37°C ± 1°C for BOL and Explant.
e. Values are not affected by temperature variation within the range 20°–43°C.
f. This parameter is automatically set to at least 85 ms f or Post-Shock Brady.
g. WhentheLVOffsetis0theLVPacefollowstheRVPaceby2.5ms
Table A-14. Atrial Tachy Parameters
Parameter Programmable Values Nominal
ATR Mode Switch
ab
Off, On On
ATR Trigger Rate
ab
( bpm)
100, 110, ..., 300
170(Toleranc5ms)
ATR Duration
ab
(cycles)
0, 8, 16, 32, 64, 128, 256, 512,
1024, 2048
8(Tolerance±1cardiaccycle)
Entry Count
ab
(cycles)
1, 2, ..., 8 8
Exit Count
ab
(cycles)
1, 2, ..., 8 8
ATR Fallback Mode
ab
VDI, DDI, VDIR, DDIR DDI
ATR Fallback Time
ab
(min:sec)
0, 0:15, 0:30, 0:45, 1:00, 1:15, 1:30,
1:45, 2:00
0:30
ATR/VTR F allback LRL
ab
( ppm)
30, 35, ..., 185
70 (Tolerance ± 5 m s)
ATR VRR
ab
Off, Min, Med, Max
Min
ATR Maximum Pacing R ate
ab
( ppm)
30, 35, ..., 185 130
ATR BiV Trigger
ab
Off, On On