User Manual Part 3

A-16
PROGRAMMABLE OPTIONS
Table A-19. 50 Hz/Manu al Burst Pacing (co n tinued)
Parameter
a
Programmable Values Nominal
Minimum Interval (ms)
20, 30, ...,750
200(Toleranc5ms)
Decrement (ms)
0, 10, ..., 50
50 (Tolerance ± 5 m s)
a. Applied to the atrium or ventricle depending on the chamber selected.
Table A- 20. Ventricular Commande d Shock
Parameter Programmable Values Nominal
Shock (stored energy) (J)
0.1, 0.3, 0.6, 0.9, 1.1, 1.7, 2, 3, 5, 6,
7, 9, 11, 14, 17, 21, 23, 26, 29, 31,
36 (HE), 41 (HE)
41 (Tolerance ± 6 0 % for 0.3 J; ±
40% for 0.6–3 J; ± 20% for 5–36
J, ± 10% for 41 J)
Coupling Interval (ms) Sync, 50, 60, ..., 500 Sync
Table A -21. VFib (Ventricular Fibrillation) I nduction
Parameter Values
VFib High
15V (nonprogrammable) (Tolerance ± 10V)
VFib Low
9V (nonprogrammable) (Tolerance ± 7V)
Table A-22. Shock on T Induction
Parameter Programmable Values Nominal
Shock (stored energy) (J)
0.1, 0.3, 0.6, 0.9, 1.1, 1.7, 2, 3, 5, 6,
7, 9, 11, 14, 17, 21, 23, 26, 29, 31,
36 (HE), 41 (HE)
1.1 J (Tolerance ± 60% for 0.3 J;
± 40% for 0.6–3 J; ± 20% for 5–36
J, ± 10% for 41 J)
Number of S1 Pulses
1, 2, ..., 30 8
S1 Interval (ms)
120, 130, ..., 750 400
Coupling Interval (ms) Sync, 10, 20, , ..., 50 0
310
Table A-23. PES (Programmed Electrical Stimulation)
Parameter
a
Programmable Values Nominal
Number of S1 Intervals (pulses)
1, 2, ..., 30 8
S2 Decrem ent
0, 10, ..., 50 0
S1 Interval (ms)
120, 130, ..., 750
600(Toleranc5ms)
S2 Interval (ms) Off, 120, 130, ..., 750 600 (Tolerance ± 5 ms)
S3 Interval (ms) Off, 120, 130, ..., 750 Off (Tolerance ± 5 ms)