Activity restrictions (if applicable)
Minimum heart rate (lower rate limit of the pulse generator)
Frequency of follow up
Travel or relocation—Follow-up arrangements should be made in advance if the patient is leaving the
country of implant
Patient ID card—the patient should be advised to carry their patient ID card at all times (a temporary
patient ID card is provided with the device, and a permanent ID card will be sent to the patient 4 to 6
weeks after the implant form is received by Boston Scientic)
Patient Handbook
The Patient Handbook is provided for each device.
It is recommended that you discuss the information in the Patient Handbook with concerned individuals both
before and after implantation so they are fully familiar with pulse generator operation.
For additional copies, contact Boston Scientic using the information on the back cover.
LEAD CONNECTIONS
Lead connections are illustrated below.
CAUTION: Prior to implantation, conrm the lead-to-pulse generator compatibility. Using incompatible leads
and pulse generators can damage the connector and/or result in potential adverse consequences, such as
undersensing of cardiac activity or failure to deliver necessary therapy.
CAUTION: If the Lead C onguration is programmed to B ipolar when a unipolar lead is implanted, pacing
will not occur.