Product Manual

SPECIFICATIONS FOR SMOKE EVACUATOR SYSTEM
Configurations Portable or Tabletop
18-hour filters, single or 3-Stage filtration in one casing
double ports 7/8" (22mm) (ULPA, Carbon)
and/or 1 1/4" (32mm)
Filters ULPA, VLSI Grade
particle size, μm 0.1 to 0.2 @ 99.999% efficiency
Filter life 18 hours
Max. Flow Setting Tubing Size
SCFM (Std. cubic ft./minute) 13/8" – NA, 1 1/4" – 38, 7/8" – 28
SLPM (Std. liters/minute) 35mm – NA, 32mm – 1076,
22mm – 793
ACFM (actual cubic feet/minute) 13/8" – NA, 1 1/4" – 59,
7/8" – 42, 1/4" – 4.8
ALPM (actual liters/minute) 35mm – NA, 32mm – 1671
22mm – 1189, 6mm – 136
Variable Flow Control Yes
Motor hp (kW) 1.4 (0.75)
Static Suction, Inches H20 (mbar) 87.1 (217)
Hoses/tubing 1/4", 7/8", 1 1/4"
6mm, 22mm, 32mm
Noise level, dBA – at min power ≤ 50
at max power ≤ 77
Footswitch Yes (Optional)
Remote control activation Yes (Optional)
Safety features Filter housing and handles
remain temperate with the
outer surface to prevent burns;
circuit breaker; CLA, CE and
UL classified/listed
Displays LED
data displayed Filter Status, Flow Rate,
Service Required
Voltage available 100/120/220/240 volts
frequency 50/60 Hz
H x W x D 10 x 6.5 x l8 inches
25.4 x 16.5 x 45.7 cm
Weight, lbs. (kg) 14.8 (6.7)
9User’s Guide • Bovie Smoke Shark
Surgical Smoke Evacuator