Instruction manual
24
WEIGHT UNIT kg
"*
CAPACITY 15
" (Input capacity)
Input capacity of scale
"*
INC.0.002/0.005
"Select interval with up arrow key [ ]↑ Select interval of the scale
"*
CAPTURE 0 ——— Read zero
"*
ADD WEIGHT 0.000
"Input the standard weight
value you want to put on scale
Input the standard weight value you want to
calibrate, at least 2/3 of capacity
"put on the standard weight on
platter
"*
DONE
SHUT DOWN
Power off scale and power on again
NOTE:
IN THE PROCESS IF GIVE THE MESSAGE “ INIT. CELL...”, WHICH INDICATE THE FAULTCALIBRATION.