Operating instructions

ER910/920 and ER910/920 AF Event Monitor
9
3. Choose program:
Choose either the DEFAULT or CUSTOM option. The DEFAULT
option only allows you to change the Contrast of the display. The
CUSTOM option allows changes to all of the following settings.
Note:
A. Factory default settings in the Monitor are indicated by a “ )( ”
in front of the setting.
B. To see the default values for the Monitor, enter the
programming mode and choose DEFAULT. The Monitor
will restart. Enter the programming mode again and
choose CUSTOM. All the values displayed will be the
defaults. There is space available on the next pages to
write the default of the Monitor.
C. Choosing the DEFAULT option will program the monitor with
the factory default values, erase previous patient data, and
restart the monitor.
Programming settings:
Display Option Description
DISPLAY
CONTRAST
Default = NA
Set the contrast of the display. A
higher number will cause darker
text on the screen.
PRE-EVENT
TIME
Default = 60 sec
Seconds of ECG data stored
before event activation time. 300
seconds total can be split between
Pre and Post events. (minimum of
5 sec)
POST-EVENT
TIME
Default = 30 sec
Seconds of ECG data stored after
event activation time. (minimum
of 5 sec)
# OF EVENTS
Default = 5
Total number of events the
Monitor will store. The maximum
number allowed is dependent on
Pre and Post Event times.
Continue on next page
ER910/920 and ER910/920 AF Event Monitor
User Defined options continued
Display Option Description
POST-EVENT
SOUND
Default = ON
ON = Sound on while recording
during the Post-Event time.
OFF = Sound off while recording
an event. See Notes about sound
for additional information.
HOURLY
REMINDER
Default = ON
ON = Ring every hour when an
event is stored in memory.
OFF = Do not remind patient an
event is stored in memory.
PREVIEW TIME
Default = 10 sec
The number of seconds the ECG
signal is displayed for each channel
when a Patient Cable is inserted.
TRANSMIT
SPEED
Default = 3X
The speed used for transmission of
data to the receiving center.
PACEMAKER
DETECTION
Default = OFF
Turn Pacemaker stimulus
detection ON/OFF
ARRHYTHMIA
DETECTION
Default = ON
Turn Arrhythmia Detection
ON/OFF. When ON, The Brady
Rate, Tachy Rate, and Pause
Duration Threshold can be set.
BRADY RATE
Default = 45 bpm
(Arrhythmia Detection must be on
to access this function)
TACHY RATE
Default = 140 bpm
(Arrhythmia Detection must be on
to access this function)
Continue on next page