Other Document

4 5
English
5. Fit the probe snuggly into the ear canal and direct towards opposite
temple. Then push and release the Start button
.
6. ExacTemp® light will pulse while temperature measurement is in progress. The light will remain
solid for 3 seconds to indicate that a successful temperature reading has been achieved.
NOTE: If the probe has been properly inserted into the ear canal during the measurement, a
long beep will sound to signal the completed measurement.
If the probe has not been constantly placed in a stable position in the ear canal, a sequence of
short beeps will sound, the ExacTemp light will go out and the display will show an error
message (POS = position error).
See «Errors and troubleshooting» section for more information.
7. The conrmation beep indicates that an accurate temperature measurement has
been taken. The result is shown on the display. Temperature reading displayed is
an equivalent of oral temperature.
To repeat
Go to step 3.
8. For the next measurement, press Eject button
to remove and discard
used Hygiene cap, and put on a new, clean Hygiene cap.
NOTE: Defaults to last age setting used if you do not change.
The Braun ThermoScan Ear thermometer turns o automatically after
60seconds of inactivity. The thermometer can also be turned o by
pressing the Power button
.
The display will briey ash OFF and it will go blank.
Temperature-taking hints
Replace disposable Hygiene cap with every use to maintain accuracy,
hygiene and to prevent cross-contamination. Earwax and dirt on the tip of
the Hygiene cap, even in a microscopic layer, can create an obstacle
between the sensor and the infrared heat emitted by the ear and impact the
accuracy of the measurement.
External factors like poor probe replacement, used Hygiene cap or dirty Hygiene
cap may inuence ear temperature measurements.
In clinical studies, precision and accuracy of ear thermometer measurement has been shown to be inuenced by
variability in user technique. Correct probe placement is essential for accurate measurements.
High fever
Fever
Normal
Red
Yellow
Green
Temperature in right and left ear diers, always compare measurements from the same ear. The natural
physiological dierence may reach 0.2 °C. In the absence of confounding factors, in high-risk population (like infants or
immunodepressed people) use the highest of the two to interpret the patient’s temperature.
In the following cases, wait 20 minutes prior to taking a temperature: extreme changes in room temperature, hearing
aid, lying on a pillow. Use the untreated ear if ear drops or other medications have been placed in the ear canal.
Memory mode
1. The last temperature taken is stored in its memory and will be
automatically displayed for 5 seconds when it is turned on again.
Night light feature
The thermometer includes a convenient night light to illuminate the display in a dim
environment.
The light will turn on when you press any button. It will stay on until the thermometer is inactive
for up to 15 seconds even after temperature is taken.
Changing the temperature scale
Your Braun ThermoScan is shipped with the Celsius (°C) temperature scale activated. If you wish to switch to Fahrenheit
(°F) and/or back from Fahrenheit to Celsius, proceed as follows:
1. Make sure the thermometer is turned o.
2. Press and hold down the Power button
. After about 3 seconds the display will show this
sequence: °C / SET / °F / SET.
3. Release the Power
button when the desired temperature scale is shown. There will be a
short beep to conrm the new setting, then the thermometer is turned o automatically.
Care and cleaning
The probe tip is the most delicate part of the thermometer. It must be clean and intact to ensure accurate
readings. If the thermometer is ever accidentally used without a Hygiene cap, clean the probe tip as follows:
Very gently wipe the surface with a cotton swab or soft cloth moistened with alcohol. After the
alcohol has completely dried out, you can put a new Hygiene cap on and take a temperature
measurement.
If the probe tip is damaged, please contact Consumer Relations.
Use a soft, dry cloth to clean the thermometer display and exterior. Do not use abrasive cleaners.
Never submerge this thermometer in water or any other liquid. Store thermometer and Hygiene
caps in a dry location free from dust and contamination and away from direct sunlight.
Additional Hygiene caps (LF 40) are available at most stores carrying Braun ThermoScan.