Vista™ basic Instructions for Use Rx Only 950787 Rev H 02/03
The CE mark confirms compliance with the “Council Directive on Medical Products 93/42/EEC” dated the 14th of June, 1993. B. Braun Melsungen AG INSPECTION ON DELIVERY Despite careful packaging, the risk of transport damage cannot be entirely excluded. Upon delivery, please check that nothing is missing. Do not use a damaged device! If the device is damaged, contact the Service Department. See “Maintenance, Storage, & Service.
Vista™ basic Contents CONTENTS Warnings & Cautions.................................................................. 3 Introduction ................................................................................ 5 Preparing the Infusion ................................................................ 7 Continuous Mode ....................................................................... 8 Options: Piggyback.......................................................................... 11 Ramp/Taper Mode ...
WARNINGS & CAUTIONS The following should be observed to avoid patient/user injury. This Instructions for Use manual contains detailed instructions and warnings on the use of the Vista™ basic Infusion Pump. Please read it completely prior to using this device. This manual is intended to reinforce the teaching given to the user by a trained health care professional or an authorized B. Braun representative. WARNINGS ♦ ♦ ♦ ♦ ♦ ♦ ♦ ♦ B.
WARNINGS & CAUTIONS ♦ ♦ ♦ ♦ After starting an infusion, make sure drops are falling in the drip chamber. If no drops are falling, make sure the roller clamp is open. If the roller clamp is open and still no drops are falling, replace and dispose of the set. The air detector cannot recognize the introduction of air at 3-way stopcocks, infusion ports, and other lines/tubes below the pump. See section entitled “Drop Sensor” for additional information on using the device without a drop sensor.
INTRODUCTION Handle for carrying Display All essential information is displayed in plain language. Green background lighting illuminates only when connected to AC power, or, in battery operation, only when a key is pressed. Keyboard Allows data entry Clear Key Resets a value to zero Alarm Light Lights up during an alarm. Display will also show flashing AAA.A.
INTRODUCTION Door Open Button Aluminum Case Easy upkeep, drip proof, and resistant to disinfectants Function Keys Allow choices of YES, NO, OK, INFO, OPTS, etc.
PREPARING THE INFUSION Prepare the Pump 1. Check for cleanliness, completeness, and damage. If unit appears damaged, do not use it. 2. Attach the pump to an IV pole by turning the knob on the pole clamp clockwise, or place the pump on a flat, stable surface. 3. Plug the pump’s power cord into an electrical outlet. Turn on the Pump 1. Press (ON/OFF). 2. Listen for 3 loud beeps. 3. Check screen for completeness of all pixels on the display. The Alarm ( ) and Infusing (6) lights will blink once.
Continuous Mode When the pump is turned on, it may display: Tubing Type VISTA Pump Set Confirm you are using a Vista™ Pump Set, then press YES. The pump may then display: Is This a New Therapy? YES NO Begins in Continuous Mode with all Returns to the beginning of the last data cleared. program with all data retained. The pump may then display: Prime Set? YES Press YES to begin priming set. • Press STOP to end priming. • Press EXIT to begin Continuous Mode. NO Begins in Continuous Mode with all data cleared.
Continuous Mode chamber. VTBD and TIME 1. Press VTBD. 2. Enter bag volume in ml. 3. Press OK. (To leave screen without saving changes, press EXIT.) 4. Press TIME. 5. Enter infusion time in hours and minutes. NOTE: For a 2 hour infusion, enter 2, 0, 0. The pump will show: Time: 02h00m 6. Press OK. (To leave screen without saving changes, press EXIT.) The pump calculates the rate and asks: Accept Rate? 7. Press OK. 8. Press to begin the infusion. Check for drops in the drip chamber. RATE and TIME 1.
Continuous Mode KEEP VEIN OPEN RATE After an infusion is complete, the pump will alarm and begin pumping at 1.0 ml/hr or the current infusion rate, whichever is less. 1. Stop the infusion. 2. Turn off the pump or reset the program. INFORMATION KEY This key is only available while the pump is infusing. 1. Press INFO. 2. The Total Infused value and the Total Time of the infusion completed are displayed. 3. Press INFO again and the Battery Left and Total Operation time of the pump is displayed. 4.
OPTIONS Piggyback The Vista™ basic contains many options to allow custom programming. To access the various options, press OPTS while the pump is stopped. Pressing OPTS each time will move you to the next option. Eventually, the OPTS key will bring you back to the mode from which you began. Depending on the model, some modes/options may have been disabled and will not appear on the display when the OPTS key is pressed. PIGGYBACK The Piggyback feature allows programming of a secondary (Piggyback) infusion.
OPTIONS Piggyback, Ramp/Taper Mode To Exit Piggyback 1. 2. 3. 4. 5. Press to stop the infusion. Press END. Press OPTS. Press OFF to turn Piggyback off Press END to return to Continuous Mode. RAMP/TAPER MODE The Ramp/Taper Mode is designed to deliver infusions, such as Total Parenteral Nutrition (TPN), which require gradual ramp up and taper down rates. The pump automatically calculates the rate changes required to match the volume, total time, and ramp/taper time parameters. 1. 2. 3. 4. 5.
OPTIONS Ramp/Taper Mode To Clear Ramp/Taper Mode 1. Press OPTS until display shows RAMP TAPER MODE. 2. Press CLR. 3. Press YES to clear all the data in Ramp/Taper Mode. 4. Press ON and continue to program a new Ramp/Taper program. To Exit Ramp/Taper Mode 1. When infusion is complete, press to stop the KVO. 2. Press EXIT. 3. Press END to return to Continuous Mode. Fast Taper The infusion can be interrupted to begin an early taper down process during the Ramp/Taper Mode Infusion. 1. 2. 3. 4. 5.
OPTIONS Program Mode PROGRAM MODE Program Mode is for infusions requiring up to 9 separate rates and volumes to be delivered one after another. This is designed to deliver infusions for stepped oncology drugs, circadian medication delivery, and other custom programs. 1. 2. 3. 4. 5. 6. 7. 8. 9. Press OPTS until display shows PROGRAM MODE. Press ON. Enter the rate of the first Step. Press OK. Enter the volume of the first Step. Press OK. Continue to enter the rates and volumes for Steps 2-9.
OPTIONS Delay A Mode DELAY A MODE The start of Continuous, Piggyback, Ramp/Taper, and Program Modes may all be delayed. The Vista™ basic also provides an option to extend the Infusion Complete alarm at the end of the infusion. 1. 2. 3. 4. 5. 6. 7. 8. 9. 10. Program the infusion. Press OPTS until display shows DELAY A MODE. Press ON. Enter in how many hours and minutes the infusion should begin. NOTE: This is NOT a clock time like 12:30 pm. Press OK.
OPTIONS Dosage Calculation DOSAGE CALCULATION This feature allows a rate to be calculated given the concentration and dose. 1. Press OPTS until display shows DOSAGE CALC. 2. Press ON. 3. If concentration unit desired is other than the one displayed, press button under the mcg? until units desired appear in the CONCENTRAT. (0*/ml) line. 4. Enter the concentration of the solution. If the concentration of the solution is not known, go to #6 below. Press C to clear current data and re-enter new data. 5.
OPTIONS Bolus Mode, Interval Bolus (Intermittent Mode) BOLUS MODE This feature allows a bolus to be given during a Continuous Mode infusion. 1. 2. 3. 4. 5. 6. 7. 8. Press OPTS until display shows BOLUS MODE. Press ON. Press RATE. Enter the rate at which the bolus should infuse, then press OK. Press END. To give a bolus during the infusion, press BOL. Enter the amount of bolus to be given in ml. Press OK. Bolus Without Entering a Volume 1.
OPTIONS Total Infused, Standby Function TOTAL INFUSED The Total Infused information screen displays the total amount of fluid in ml that has been pumped since the last time the Total Infused was cleared. To Clear the Total Infused 1. Press OPTS until display shows Tot Infused=. 2. Press . 3. Press END. STANDBY This function allows the pump to be put on standby while retaining all set infusion data. This function is only available in Continuous Mode. 1. 2. 3. 4.
OPTIONS Occlusion Pressure, Drop Sensor, Battery OCCLUSION PRESSURE The pressure at which a downstream occlusion alarm will be triggered may be varied. The higher the pressure limit, the less sensitive the pump is to changes in fluid resistance. 1. Press OPTS until display shows OCCLUSION PRESSURE. 2. Press (--) to change pressure to Low. Press (+) to change pressure to High. 3. Press END to return to Continuous Mode. NOTE: Low corresponds to approximately 8 PSI and high to approximately 17 PSI.
OPTIONS Data Lock, Contrast Selection DATA LOCK This features prevents unauthorized use by locking out all keys except ,and , . 1. Press OPTS until display shows DATA LOCK. 2. Press ON. 3. Press END to return to Continuous Mode. To Turn Off Data Lock 1. Press OPTS until display shows DATA LOCK. 2. Press the decimal key. 3. Press OFF. 4. Press END to return to Continuous Mode. CONTRAST SELECTION This feature allows the contrast of the screen to be changed for best possible viewing. 1.
OPTIONS Adjust Clock, Loudness ADJUST CLOCK This feature allows the time and date to be changed. 1. Press OPTS until display shows ADJUST CLOCK. 2. Press DAT to enter the date. (Press C to clear the current data and reenter new data.) 3. Enter the date using two digits for month, day and year. Example: November 5, 2001 is entered as 11 05 01 4. Press OK 5. Press TIME to enter the time of day. Note: The time is set using 24 hour Military Time. Example: 6:30pm is entered as 18:30 6. Press OK 7.
OPTIONS Alarms In alarm situations, press the infusion by pressing Alarm to silence the alarm. Correct the problem. Resume .
AC, 12 V, BATTERY, & CC INTERFACE AC and 12 V ♦ Check AC voltage as per rating label. ♦ Insert/screw on power lead (12 V lead in ambulance car) on back. Battery Operation ♦ Switching to internal battery is automatic in the event of power failure or if 12 V or AC is not connected. ♦ Charge battery on initial use, following a battery alarm, or after 2 months of non-use. ♦ Charging time: approx. 16 hours. Longer charging has no detrimental effect. ♦ Full battery provides operation for 3.5 hours at 125 mL/hr.
MAINTENANCE, STORAGE, & SERVICE Maintenance Clean the pump with a soft, lint-free cloth or swab dampened with soap and water, a general non staining chemical disinfectant, isopropyl alcohol (15%), derivatives of phenylphenol (5%) or ammonium chloride (10%). As a general recommendation for cleaning and disinfecting this device, use a lint-free cloth dampened (not soaked) with a 0.5% bleach (Sodium Hypochlorite) solution (10% solution with water).
SPECIFICATIONS Mechanism Dimensions WxHxD Weight Power Requirements Leakage Current Battery Battery Operating Time Rate Range Volume to be Infused Range Occlusion Pressure KVO Rate Flow Rate Accuracy Alarms Air-in-Line Alarm Memory Tubing Modes/Features Operating Temperatures ™ Vista basic Instructions for Use 950787 Rev H 02/03 Linear peristaltic volumetric infusion pump 5.5 in (14cm) x 9.4 in (24cm) x 7.8 in (20cm) 6.8 lbs. (3.1 kg) 100-120 VAC, 50-60 Hz, 13 Watts Max 30µA 7.2V, 1.
B. Braun Medical Inc. 824 Twelfth Ave. Bethlehem, PA 18018 ©2001 B. Braun Medical Inc. ™ Vista basic Instructions for Use 950787 Rev H 02/03 Tel 800-227-2862 www.bbraunusa.