PS-500 Compact Light Weight Floor Scale User Instructions AWT35-501194 Issue AA June 2013
© Brecknell, LLC 2013. All rights reserved. No part of this publication may be reproduced, stored in an electronic retrieval system, or transmitted in any form or by any means, electronic, mechanical, photocopying, recording or otherwise without the prior written consent of the copyright owner, or as permitted by law or under license. Full acknowledgment of the source must be given. Brecknell is a registered trade mark of the Brecknell, LLC.
Table of Contents Chapter 1 General Information and Warnings ........................................................................................ 3 About this Manual .............................................................................................................. 3 Text Conventions ........................................................................................................ 3 Special Messages ........................................................................................
Commands and Response .............................................................................................. 30 Chapter 7 Troubleshooting ..................................................................................................................... 32 Display Characters ........................................................................................................... 32 Display Symbols ...................................................................................................
1 General Information and Warnings 1.1 About this Manual This manual is divided into chapters by the chapter number and the large text at the top of a page. Subsections are labeled as shown by the 1 and 1.1 headings shown above. The names of the chapter and the next subsection level appear at the top of alternating pages of the manual to remind you of where you are in the manual. The manual name and page numbers appear at the bottom of the pages. 1.1.
1.2 Warnings l Read all operating instructions carefully before use. l Avoid lengthy exposure to extreme heat or cold. Your scale works best when operated at normal room temperature. Always allow the scale to acclimate to a normal room temperature before use. l Allow sufficient warm up time. Turn the scale on and wait for a few minutes if possible. This will give the internal components a chance to stabilize before weighing. l These electronic scales are precision instruments.
Canada This digital apparatus does not exceed the Class A limits for radio noise emissions from digital apparatus set out in the Radio Interference Regulations of the Canadian Department of Communications. Le présent appareil numérique n’émet pas de bruits radioélectriques dépassant les limites applicables aux appareils numériques de la Classe A prescrites dans le Règlement sur le brouillage radioélectrique edicté par le ministère des Communications du Canada. 1.
2 Specifications 2.1 Scale Platform 2.1.1 Capacity l 250 kg x 0.1 kg l 500 lb x 0.2 lb l 500 lb x 5 oz l 22” L x 22” W x 2” H 2.1.2 Dimension m l 36” L x 22” W x 2” H m l 559 mm (L) x 559 mm (W) x 51 mm (H) 906 mm (L) x 559 mm (W) x 51 mm (H) 42” L x 22” W x 2” H m 1067 mm (L) x 559 mm (W) x 51 mm (H) 2.1.3 Options 2.2 l Table top stand l Floor stand Scale Indicator l Input signal range: 0mV - +30mV l Sensitivity: >0.
2.2.2 Dimension 6.5” L x 3.2” H x 1.2” W l m 164 mm (L) x 762 mm (W) x 64 mm (H) 2.2.3 Display l 5-digit,7-segment, 0.625" (16mm) LCD l 4 push button keys 2.2.4 Keypad 2.2.5 Environment Working temperature 5° to 35°C Storage temperature -10°C to 70°C Humidity ≤95% RH without condensation 2.2.6 Load Cell Excitation Because more than one load cell can be used on a scale, the following are required for the load cell set to be used with this indicator. Voltage 4.4VDC Max.
3 Introduction This chapter introduces indicator mounting, display and keypad operation in normal weigh mode. 3.1 Indicator Mounting The PS-500 is supplied with an ABS plastic bracket. Wall mount vertically or bench mount horizontally depending on your needs. 3.2 Front Panel The front panel incorporates the display and keypad.
3.3 Display The PS-500 indicator utilizes a 7 segment LCD (Liquid Crystal Display) providing the weight and system information. Arrows will illuminate to designate the current mode. Max lb kg d= oz Figure 3.1 PS-500 Display Table 3.1 LCD Display Annunciators and Definitions Display Symbol Description ZERO The scale is at zero point and the gross weight is 0. HOLD The scale is in HOLD mode. The current live weight is being displayed when the arrow flashes.
3.4 Keyboard The keyboard consists of four keys, some of which have multiple functions. Figure 3.2 SBI-521 Keypad Table 3.2 Function of the Keys (Normal Working Mode) Key Function If this key is only set for HOLD (P2=0), press to enter or exit the HOLD mode. If this key is only set for PRINT (P2=1), press to output the data according to P4 setting. If this key is only set for HOLD and PRINT (P2=2), press to output data according to P4 setting and hold the current weight.
3.5 Navigating the Weigh Mode 3.5.1 Power the Indicator Turn the indicator on or off with the [ON/OFF] key. 3.5.2 Enter or Exit the HOLD Mode Press the [HOLD] key. 3.5.3 Zero When the weight is stable and within the zero range, press the [ZERO] key to set a new zero point. Refer to Table 3.3 for zero limitations. 3.5.4 Tare When the gross weight is larger than zero and the scale is stable, press the [TARE] key.
key function Standard Weight on platform Data in TARE memory unit Tare key Zero key Europe ≤0 no No action Zero yes Clear the tared weight Zero and clear the tared weight no Tare Zero >0 yes None (same with Europe) ≤0 >0 Zero and clear the tared weight no No action Zero yes Clear the tared weight Zero and clear the tared weight no Tare Zero yes Zero and clear the tared weight 3.5.6 Output Data (Print) When scale is stable press the [PRINT] key. 3.5.
Display Division Value Calibration Division Value kg lb lb:oz 0.005kg 0.005kg 0.01 lb 0.2 oz 0.05kg 0.05kg 0.1 lb 2oz 0.5kg 0.5kg 1 lb Not available 5kg 5kg 10 lb Not available 50kg 50kg Not available Not available Table 3.5 Use lb as Primary Unit PS500 User Instructions Display Division Value Calibration Division Value kg lb lb:oz 0.0001lb Not available 0.0001lb Not available 0.001 lb 0.0005 kg 0.001 lb Not available 0.01 lb 0.005 kg 0.01 lb 0.2 oz 0.1 lb 0.
3.6 Hold Function HOLD function can be used to freeze a displayed number. In this mode, the scale can capture a dynamic number, hold a stable number, or average a unstable number and then HOLD (freeze) this number temporarily for the user to view or record. The HOLD function can be used in normal weighing mode, counting mode and percent weighing mode. After entering HOLD mode, the speed of A/D converter can be increased to 80Hz (if USER-HOLD-AD.H.
4 Setup Mode To set up the indicator, you must first enter the appropriate menu mode. The front panel keys become directional navigators to move around in the menus. See Table 4.1 for details. 4.1 4.2 Entering the Setup Menu 1. Press and hold the [ON/OFF/ZERO] and [UNIT] keys unit setup is displayed. 1a. This indicator offers 19 different setup parameters. Navigating the Setup Menu 1.
4.3 Setup Menu Parameters This section provides more detailed descriptions of the selections found in the Setup Menu. The menu table shows the options and default parameter in LCD display format to coinside with the actual display. Table 4.2 Parameter Choices and Explanations SubMenu1 Option Default P1 00-15 05 p2 0-2 p3 0-2 Parameter Description Auto off time 00 = no auto off 01-15 = auto off time in minutes.
SubMenu1 Option Default P5 0-4 3 Baud rate for RS-232 0 = 1200 bps 1 = 2400 bps 2 = 4800 bps 3 = 9600 bps 4 = 19200 bps p6 0-2 0 RS-232 protocol 0 = 8NO 1 = 7O1 2 = 7E1 p7 00-31 9 Scale resolution select Refer to Table 4.3 for resolution choices and values. P8 0-2 1 Calibration division size Refer to Table 4.4 and Table 4.5 0=1 1=2 2=5 p9 0-5 1 Decimal point in calibration Refer to Table 4.4 and Table 4.5 0=1 1 = 0.1 2 = 0.01 3 = 0.001 4 = 0.
18 SubMenu1 Option Default Parameter Description Comment p15 0-3 1 Weight signal not within power on zero point range 0 = current weight 1 = calibration zero-point 2 = switch-off zero-point 3 = continuously display 0----- p16 0-8 8 Zero tracking range Choose the zero tracking range as per the stability of weighing system accuracy. 0 = 0d, no tracking 1 = +0.25d 2 = +0.5d 3 = +1d 4 = +1.
Table 4.3 Calibration Resolution (P7) Parameter Choice Calibration Resolution Parameter Choice Calibration Resolution 00 500 16 7500 01 600 17 8000 02 750 18 10000 03 800 19 12000 04 1000 20 15000 05 1200 21 20000 06 1500 22 25000 07 2000 23 30000 08 2400 24 35000 09 2500 25 40000 10 3000 26 50000 11 3500 27 60000 12 4000 28 70000 13 5000 29 75000 14 6000 30 80000 15 7000 31 100000 Table 4.
Display Division Value Calibration Division Value kg lb lb:oz 0.005kg 0.005kg 0.01 lb 0.2 oz 0.05kg 0.05kg 0.1 lb 2oz 0.5kg 0.5kg 1 lb Not available 5kg 5kg 10 lb Not available 50kg 50kg Not available Not available Table 4.5 Calibration Units (lb) per P8, P9 and P10 20 Display Division Value Calibration Division Value kg lb lb:oz 0.0001lb Not available 0.0001lb Not available 0.001 lb 0.0005 kg 0.001 lb Not available 0.01 lb 0.005 kg 0.01 lb 0.2 oz 0.1 lb 0.
4.4 Relationship of Capacity and P7, P8 and P9 Settings Table 4.6 Capacity Unit is kg or lb (Count by 1) Division set by P8 (P8 = 0) and P9 Resolution set by P7 0.0001 0.001 0.01 0.1 1 10 500 0.0500 0.500 5.00 50.0 500 5000 600 0.0600 0.600 6.00 60.0 600 6000 750 0.0750 0.750 7.50 75.0 750 7500 800 0.0800 0.800 8.00 80.0 800 8000 1000 0.1000 1.000 10.00 100.0 1000 10000 1200 0.1200 1.200 12.00 120.0 1200 12000 1500 0.1500 1.500 15.00 150.
Table 4.7 Capacity Unit is kg or lb (Count by 2) 22 Division set by P8 (P8 = 1) and P9 Resolution set by P7 0.0002 0.002 0.02 0.2 2 20 500 0.1000 1.000 10.00 100.0 1000 10000 600 0.1200 1.200 12.00 120.0 1200 12000 750 0.1500 1.500 15.00 150.0 1500 15000 800 0.1600 1.600 16.00 160.0 1600 16000 1000 0.2000 2.000 20.00 200.0 2000 20000 1200 0.2400 2.400 24.00 240.0 2400 24000 1500 0.3000 3.000 30.00 300.0 3000 30000 2000 0.4000 4.000 40.00 400.
Table 4.8 Capacity Unit is kg or lb (Count by 5) Division set by P8 (P8 = 2) and P9 Resolution set by P7 0.0005 0.005 0.05 0.5 5 50 500 0.2500 2.5000 25.00 250.0 2500 25000 600 0.3000 3.000 30.00 300.0 3000 30000 750 0.3750 3.750 37.50 375.0 3750 37500 800 0.4000 4.000 40.00 400.0 4000 40000 1000 0.5000 5.000 50.00 500.0 5000 50000 1200 0.6000 6.000 60.00 600.0 6000 60000 1500 0.7500 7.500 75.00 750.0 7500 75000 2000 1.0000 10.000 100.00 1000.
4.5 24 Exit the Setup Menu 1. Power off the indicator by pressing and holding the [ZERO/ON/OFF] key. 2. Move the slide switch on the rear cover back to the left and replace the metal protective plate. Refer to section 4.2 for location. 3. Turn the indicator back on by pressing the [ON] key. The display will go through a digit check, then settle into Normal Operating mode. All front panel keys will now return to their normal mode of operation.
5 Calibration The config/cal switch must be set to the ON position in order to calibrate the indicator. NOTE: More than 25% of the full scale weight is needed for calibration. 5.1 Calibration Mode Before calibration, prepare the standard weight (more than 25% of full scale). 1. Remove any weight on the scale. 2. When in normal weighing mode, press and hold down [TARE] and [ON/OFF/ ZERO] keys to enter calibration mode. 3. When the indicator displays cal-, the scale is ready for calibration.
9. Cal.p1 will be displayed. This is the second calibration point. The default standard weight is 50% of full scale. Load weight from 12.5% to 100% of full scale. 10. Use the [PRINT/HOLD] and [UNIT] keys to input the loaded weight. 10a. If the input data is larger than 199999 it will be shown by first four digits and last four digits: Hxxx and Lxxx. 10b.
In this mode, the proper ADC data at zero point can be viewed by examining the A/D data for loaded weight. m If the ADC increase for full capacity is NFS, the power-on zero range is set to Zp% FS (P12 setting) and zero key range is set to Zk% FS (P13 setting). Then proper ADC data of zero point is larger than (Zp%+ Zk%) x NFS. 1. From the weigh mode press and hold the [ON/OFF/ZERO] and [HOLD/PRINT] key until Code is shown. The indicator will show the firmware version. 2.
6 Serial Communication 6.1 Com Port 1 COM1 is a RS-232 bi-directional port. Communication wires are connected to the RS232 connector using TXD1, RXD1 and GND. Table 6.
Table 6.2 Symbols Used Line Feed character (hex 0AH) Carriage Return character (hex 0DH) End of Text character (hex 03H) Space (hex 20H) H1H2H3 Four current status bytes Polarity character: "?" or " " W1---W7 Reading data, 1-7 bytes (seven digits) Decimal point U1U2 Measure units, kg, lb, lb:oz, % or pcs; 2-5 bytes Table 6.
6.4 Commands and Response Command: W (57h 0dh), request current reading Response ^^^^^^^^U1U2 H1H2H3 ---over capacity _ _ _ _ _ _ _ _ U1U2 H1H2H3 ---under capacity - - - - - - - - U1U2 H1H2H3 ---zero-point error Note: If the weigh unit is lb:oz, U1U2 oz in above responses W1W2W3W4W5W6W7 U1U2 H1H2H3 ---scale is stable and the current weigh unit is kg or lb.
Command: U (55h 0dh) Response Changes units of measure (simulate UNIT key) and return scale status with new units, The new measure unit should be allowed to use U1U2 H1H2H3 Command: L (4ch 0dh) Response If Hold function can be activated, it will enable/disable hold function (simulate HOLD key) and returns scale status. H1H2H3 Command: X (58h 0dh) Response Power off the scale, just like press down the ON/OFF key to turn off the scale.
7 Troubleshooting This chapter gives explanations on commonly seen errors, display characters and display symbols. 7.
7.2 Display Symbols Symbol 7.3 Description 0------ Zero is over the setting range 0_______ Zero point is below the setting range Ad------ Signal to ADC is over maximum range Ad______ Signal to ADC is below minimum range -------- Weight is over upper limitation or display data is over limitation ________ Weight is below lower limitation eep.e0 EEPROM can’t be accessed. EEP.E1 The parameters are not the same as the backup data EEP.E2 The setting parameter(s) is not in normal range Lo.
34 PS500 User Instructions
Brecknell USA 1000 Armstrong Dr. Fairmont MN 56031 Tel:507-238-8702 Fax:507-238-8271 Email: sales@brecknellscales.com http://www.brecknellscales.com Brecknell UK Foundry Lane, Smethwick, West Midlands, England B66 2LP Tel:+44 (0) 8452 46 6717 Fax:+44 (0) 8452 46 6718 Email: sales@brecknellscales.co.uk http://www.brecknellscales.
PS-500 Balance de plancher légère et compacte Mode d’emploi AWT35-501194 Numéro AA Juin 2013
© Brecknell, s.e.c. 2013. Tous droits réservés. Le présent document ne peut être, en tout ou en partie, reproduit, enregistré dans un système d’extraction électronique, ou transmis de quelque manière ou sous quelque forme que ce soit, électroniquement, mécaniquement, par photocopie, enregistrement ou autre, sans le consentement écrit préalable du détenteur du droit d'auteur, sous réserve des dispositions d’une loi ou d’une licence. La source doit être totalement reconnue.
Table des matières Chapter 1 Renseignements et avertissements généraux ...................................................................... 3 À propos de ce manuel ...................................................................................................... 3 Conventions d’écriture ................................................................................................. 3 Messages spéciaux ....................................................................................................
Commandes et réponses ................................................................................................. 30 Chapter 7 Dépannage .............................................................................................................................. 32 Caractères d’affichage ..................................................................................................... 32 Symboles affichés .............................................................................................
1 Renseignements et avertissements généraux 1.1 À propos de ce manuel Il est divisé en chapitres; chaque chapitre est numéroté et est identifié par les gros caractères en haut de page. Les subdivisions sont identifiées tel qu’indiqué par les entêtes 1 et 1.1 indiquées ci-dessus. Le nom du chapitre et le niveau de subdivision suivant sont indiqués en haut de page, à toutes les deux pages du manuel, pour vous rappeler où vous êtes, dans celui-ci.
1.2 Avertissements l Veuillez lire attentivement l’ensemble du mode d’emploi avant de vous en servir. l Éviter l'exposition prolongée à la chaleur ou au froid extrême. Votre balance fonctionne mieux à la température ambiante normale. Toujours permettre à la balance de s’habituer à la température ambiante avant de vous en servir. l Prévoir un réchauffement adéquat. Allumer la balance et attendre quelques minutes, dans la mesure du possible.
États-Unis Cet équipement a été testé et déclaré conforme aux restrictions des appareils numériques de catégorie A, conformément à la section 15 des règles de la FCC. Ces restrictions sont conçues pour offrir une protection raisonnable contre les interférences nocives lorsque l’équipement est utilisé dans un environnement commercial.
2 Caractéristiques 2.1 Plateforme de la balance 2.1.1 Capacité l 250 kg x 0,1 kg l 500 lb x 0,2 lb l 500 lb x 5 oz l 559 mm (L) x 559 mm (L) x 51 mm (H) 2.1.2 Dimensions m l 906 mm (L) x 559 mm (L) x 51 mm (H) m l 22 po L x 22 po L x 2 po H 36 po L x 22 po L x 2 po H 1 067 mm (L) x 559 mm (L) x 51 mm (H) m 42 po L x 22 po L x 2 po H 2.1.3 Options 2.
2.2.2 Dimensions 164 mm (L) x 762 mm (L) x 64 mm (H) l m 6,5 po L x 3,2 po H x 1,2 po L 2.2.3 Affichage l ACL 5 chiffres, 7 segments, 16 mm (0,625 po.) l 4 touches à pression 2.2.4 Clavier 2.2.5 Environnement Température de fonctionnement De 5 ° à 35 °C Température d’entreposage De –10 °C à 70 °C Humidité ≤95 % RH sans condensation 2.2.
3 Introduction Ce chapitre présente le montage de l’indicateur, l’affichage et le fonctionnement du clavier en mode de pesée normale. 3.1 Montage de l’indicateur Le PS-500 est fourni avec une pince de plastique ABS. Monter verticalement au mur ou horizontalement sur un support selon vos besoins. 3.2 Panneau avant Le clavier incorpore l’affichage et le clavier.
3.3 Affichage L'indicateur du PS-500 utilise un ACL (affichage à cristaux liquides) à 7 segments affichant le poids et des données système. Les flèches s’illumineront pour indiquer le mode en cours. d= Max lb kg oz Figure 3.1 Affichage du PS-500 Table 3.1 Voyants d’affichage ACL et définitions Symbole affiché ZERO HOLD (RETENIR) NET Description La balance est au point zéro et le poids brut est à 0. La balance est en mode HOLD (RETENIR). Le poids actuel est affiché lorsque la flèche clignote.
3.4 Clavier Le clavier est constitué de quatre touches dont certaines ont plus d'une fonction. Figure 3.2 Clavier SBI-521 Table 3.2 Fonction des touches (en mode normal) Touche Fonction Si cette touche n’est définie que pour le mode HOLD (RETENIR) (P2=0), appuyer pour entrer ou quitter le mode HOLD (RETENIR). Si cette touche n’est définie que pour le mode PRINT (IMPRIMER) (P2=1), appuyer pour exporter les données selon le paramètre P4.
3.5 Parcourir le mode de pesée 3.5.1 Allumer le témoin Allume ou éteint le témoin avec la touche [ON/OFF]. 3.5.2 Entrer ou quitter le mode HOLD (RETENIR) Appuyer sur la touche [HOLD] (RETENIR). 3.5.3 Zéro Lorsque le poids est stable et dans l'intervalle du zéro, appuyez sur la touche [ZERO] pour fixer un nouveau point zéro. Pour connaître les restrictions quant au tarage, consulter Table 3.3. 3.5.
Standard Poids sur le plateau Données dans l’unité de mémoire TARE (TARE) Europe ≤0 >0 principales fonctions Touche de tare Touche zéro non Aucune action Zéro oui Supprimer le poids à vide Annuler et supprimer le poids à vide non Tare Zéro oui Aucun (identique à l’Europe) ≤0 >0 Annuler et supprimer le poids à vide non Aucune action Zéro oui Supprimer le poids à vide Annuler et supprimer le poids à vide non Tare Zéro oui Annuler et supprimer le poids à vide 3.5.
Valeur de division du calibrage Afficher la valeur de division kg lb lb:oz 0,005 kg 0,005 kg 0,01 lb 0,2 oz 0,05 kg 0,05 kg 0,1 lb 2 oz 0,5 kg 0,5 kg 1 lb Non disponible 5 kg 5 kg 10 lb Non disponible 50 kg 50 kg Non disponible Non disponible Table 3.
3.6 Fonction Hold (Retenir) La fonction HOLD (RETENIR) peut servir à bloquer un nombre affiché. Dans ce mode, la balance peut capturer une variable, une constante, ou la moyenne d'une variable et mettre ce chiffre en mode HOLD (RETENIR) temporairement pour permettre à l’utilisateur de le consulter ou de l’enregistrer. La fonction HOLD (RETENIR) peut être utilisée en mode de pesée normale, en mode de comptage, et en mode de pourcentage de poids.
4 Mode de configuration Afin de configurer l’indicateur, vous devez d'abord activer le mode approprié. Les touches du clavier deviennent des flèches permettant de se déplacer dans les menus. Pour de plus amples renseignements, voir le Table 4.1. 4.1 4.2 Entrer dans le menu de configuration 1. Tenir enfoncée les touches [ON/OFF/ZERO] et [UNIT] (UNITÉ) jusqu’à ce que setUp (configuration) soit affiché. 1a. Cet indicateur propose 19 paramètres différents. Parcourir le menu Setup (Configuration) 1.
4.3 Paramètres du menu de configuration La présente section décrit plus en détails les choix offerts dans le menu de configuration. Le tableau du menu présente les options et le paramètre par défaut en format d'affichage ACL pour reprendre l'affichage réel. Table 4.2 Choix de paramètre et explications SubMenu1 Option Par défaut Description du paramètre P1 00-15 05 Délai d’arrêt automatique 00 = fonction désactivée 01–15 = délai d’arrêt automatique en minutes.
SubMenu1 Option Par défaut P5 0-4 3 Vitesse de transfert du RS-232 0 = 1 200 bps 1 = 2 400 bps 2 = 4 800 bps 3 = 9 600 bps 4 = 19 200 bps p6 0-2 0 Protocole RS-232 0 = 8NO 1 = 7O1 2 = 7E1 p7 00-31 9 Choix de la résolution de la Le Table 4.3 présente les choix de résolution et balance les valeurs. P8 0-2 1 Format de division du calibrage Voir Table 4.4 et Table 4.5 0=1 1=2 2=5 p9 0-5 1 Point décimal au calibrage Voir Table 4.4 et Table 4.
18 SubMenu1 Option Par défaut Description du paramètre Commentaire p15 0-3 1 Signal de poids hors de l’intervalle zéro au démarrage 0 = poids actuel 1 = calibrage du point zéro 2 = annulation du point zéro 3 = affichage continu de 0----- p16 0-8 8 Intervalle de suivi du zéro Choisir l'intervalle de suivi du zéro selon la stabilité de la précision du système de pesée.
Table 4.3 Résolution du calibrage (P7) Choix de paramètre Résolution du calibrage Choix de paramètre Résolution du calibrage 00 500 16 7 500 01 600 17 8 000 02 750 18 10 000 03 800 19 12 000 04 1 000 20 15 000 05 1 200 21 20 000 06 1 500 22 25 000 07 2 000 23 30 000 08 2 400 24 35 000 09 2 500 25 40 000 10 3 000 26 50 000 11 3 500 27 60 000 12 4 000 28 70 000 13 5 000 29 75 000 14 6 000 30 80 000 15 7 000 31 100 000 Table 4.
Valeur de division du calibrage Afficher la valeur de division kg lb lb:oz 0,005 kg 0,005 kg 0,01 lb 0,2 oz 0,05 kg 0,05 kg 0,1 lb 2 oz 0,5 kg 0,5 kg 1 lb Non disponible 5 kg 5 kg 10 lb Non disponible 50 kg 50 kg Non disponible Non disponible Table 4.
4.4 Relation de capacité et paramètres P7, P8 et P9 Table 4.
Table 4.
Table 4.
4.5 24 Quitter le menu Setup (Configuration) 1. Éteindre l'indicateur en tenant enfoncé la touche [ZERO/ON/OFF]. 2. Déplacer le commutateur sur le couvercle arrière vers la gauche et remplacer la plaque de protection en métal. Pour savoir où il se trouve, voir la section 4.2. 3. Rallumer le témoin en appuyant sur la touche [ON]. L’affichage effectuera une vérification des chiffres, puis passera au mode de fonctionnement normal.
5 Calibration Le commutateur config/cal doit être en position ON (MARCHE) afin de calibrer l’indicateur. NOTA : Le calibrage doit être fait avec un poids de plus de 25 % de la capacité de la balance. 5.1 Mode da calibrage Avant le calibrage, préparer le poids standard (plus de 25 % de la capacité de la balance). 1. Retirer tout poids de la balance. 2. En mode de pesée normale, tenir enfoncées les touches [TARE] (TARE) et [ON/OFF/ZERO] pour entrer en mode de calibrage. 3.
9. CAL.p1 sera affiché. Il s’agit du deuxième point de calibrage. Le poids standard par défaut est de 50 % de la capacité de la balance. Poids de charge de 12,5 % à 100 % de la capacité de la balance. 10. À l’aide des touches [PRINT/HOLD] (IMPRIMER/RETENIR) et [UNIT] (UNITÉ), entrer le poids chargé. 10a. Si les données saisies sont plus élevées que 199 999, seuls les 4 chiffres des extrémités seront indiqués : Hxxx et Lxxx. 10b.
Dans ce mode, les données ADC adéquates au point zéro peuvent être consultées en examinant les données A-D pour le poids chargé. m Si l’augmentation ADC de la capacité totale est NFS, l’intervalle zéro au démarrage est fixé à Zp % FS (réglage P12) et l’intervalle de la touche zéro est fixé à Zk % FS (réglage P13). C’est que la valeur ADC adéquate du point zéro est supérieure à (Zp %+ Zk %) x NFS. 1.
6 Communications de série 6.1 Port Com 1 COM1 est un port bidirectionnel RS-232. Les fils de communication sont branchés au connecteur RS-232 utilisant TXD1, RXD1 et GND. Table 6.
Table 6.2 Symboles utilisés Caractère d’alimentation (hex 0AH) Caractère de retour de chariot (hex 0DH) Caractère de fin de texte (hex 03H) Espace (hex 20H) H1H2H3 Quatre octets de l’état actuel Caractère de polarité : « ? » ou « » W1---W7 Données de lecture, 1-7 octets (sept caractères) Point décimal U1U2 Unités de mesure units, kg, lb, lb:oz, % ou pcs; 2–5 octets Table 6.
6.4 Commandes et réponses Commande : W (57h 0dh), demander la lecture actuelle Réponse ^^^^^^^^U1U2 H1H2H3 ---au-dessus de la capacité _ _ _ _ _ _ _ _ U1U2 H1H2H3 ---sous la capacité - - - - - - - - U1U2 H1H2H3 ---erreur de point zéro Nota : Si l’unité de pesée est lb:oz, U1U2 oz dans les réponses ci-dessus W1W2W3W4W5W6W7 U1U2 H1H2H3 ---la balance est stable et l’unité de pesée actuelle est kg ou lb.
Commande : U (55h 0dh) Réponse Change l’unité de mesure (simule la touche UNIT (UNITÉ)) et retour à l’état actuel de la balance, avec la nouvelle unité. La nouvelle unité de mesure devrait être disponible. U1U2 H1H2H3 Commande : L (4ch 0dh) Réponse Si la fonction Hold (Retenir) peut être activée, elle activera ou désactivera la fonction Hold (simulation de la touche HOLD (RETENIR)) et retour à l’état de la balance.
7 Dépannage Ce chapitre fournit des explications sur les erreurs fréquentes, les caractères et les symboles affichés. 7.
7.2 Symboles affichés Symbole 7.3 Description 0------ Le zéro est au-dessus de l’intervalle du paramètre 0_______ Le point zéro est sous l’intervalle du paramètre Ad------ Le signal à ADC est au-delà de l’intervalle Ad______ Le signal à ADC est en-deçà de l'intervalle -------- Le poids dépasse la limite maximale ou les données d’affichage dépassent les limites ________ Le poids est inférieur à la limite minimale eep.e0 EEPROM est inaccessible EEP.