Owner manual

Brookfield Engineering Laboratories, Inc. Page 14 Manual No. M/98-211-A0701
Time
XXX.X
M
S
T
Model
S
p
indle
Tor
q
ue
Viscosit
y
Shear Stress
cP or mPas
D/CM2 or
N/M2 =
D/CM2/10
XXXXX
XX
X.XX
XXX.X
XXXXX
XX.X
0 <= cP <= 9.99
10 <= cP <= 999.9
1000 <= cP <= 99999
00 <= S <= 99
-10.1 <= T <= 99.9
See Model Table D2 - A
pp
D
12345
RV
31
82.4
3.16
123.8
XXXeX 100000 <= cP <= 51200000000 123e3 to 8
4.56
12345
0 <= D/CM2 <= 9.99X.XX
XXX.X
XXXXX
10 <= D/CM2 <= 999.9
1000 <= D/CM2 <= 99999
234.5
Shear Rate 1/SEC
Tem
p
erature T
-99.9 <= T <= 300.0 -10.3
X.XXX 0 <= 1/SEC <= 9.999
XX.X
XXXXX
10 <= 1/SEC <= 99.9
100 <= 1/SEC <= 99999
1.234
20.7
200
0.01 <= RPM <= 0.99
RPMRPM
Item Print Format ExampleRange
0.09X.XX
XX.X 0.1 <= RPM <= 99.9 2.4
XXX.X 100 <= RPM <= 250 150.0
XXXeX 100000 <= D/CM2 <= 999999 123e3
Z
XX:XX 00:00 <= Z <= 99.59 05:32
Time
Table 1
OUT OF RANGE INDICATORS
The DV-III+ is capable of measuring instrument torque within the range of 0 to 100%. Based on
this measurment, viscosity and shear stress are calculated. Brookfield recommends that data be
collected only in the range of 10 to 100%. Any data collected outside of this range is considered
invalid.
The DV-III+ provides the following display indicators when the measurement point is outside of
the 10-100% acceptable range.
TORQUE GREATER THAN 100%
When Rheometer torque exceeds 100%, the parameter display field will show “EEEE” for torque,
viscosity and shear stress.
Figure 8
TORQUE LESS THAN 10%
When Rheometer torque drops below ten (10) percent, the Rheometer will continue to display
measurement (%, cP, D/cm
2
) values with units flashing: