Owner manual

Brookfield Engineering Laboratories, Inc. Page 57 Manual No. M/98-211-A0701
Universal Spindle Ranges
Example 1: Determine the full scale viscosity range of the LV3 spindle running on an RV series
Rheometer at 45 RPM.
Spindle Speed = 45 RPM
LV Spindle Range Coefficient for RV series Rheometer = 1,280,000
Full Scale Viscosity Range = 1,280,000 = 28,444 cP (mPas)
45
Example 2: Determine the full scale viscosity range of the LV3 spindle running on a 2xHA
Rheometer at 103.5 RPM.
Spindle Speed = 103.5 RPM
LV3 Spindle Range Cooefficient for 2xHA Rheometer = 2 x 2,560,000
Full Scale Viscosity Rnage = 5,120,000 = 49,468 cP (mPas)
103.5
Note: The maximum viscosity taht should be taken from the DV-III+ Rheometer is at 100%
of any full scale spindle/speed range.
DIN-81 81 3,420 36,500 73,000 292,000
DIN-82 82 3,420 36,500 73,000 292,000
DIN-83 83 11,340 121,300 242,600 970,400
DIN-85 85 1,144 12,200 24,400 97,600
DIN-86 86 3,420 36,500 73,000 292,000
SC4-14 14 117,200 1,250,000 2,500,000 10,000,000
SC4-15 15 46,880 500,000 1,000,000 4,000,000
SC4-16 16 120,000 1,280,000 2,560,000 10,240,000
SC4-18 18 3,000 32,000 64,000 256,000
SC4-21 21 4,688 50,000 100,000 400,000
SC4-25 25 480,000 5,120,000 10,240,000 40,960,000
SC4-27 27 23,4000 250,000 500,000 2,000,000
SC4-28 28 46,8880 500,000 1,000,000 4,000,000
SC4-29 29 93,750 1,000,000 2,000,000 8,000,000
SC4-31 31 30,000 320,000 640,000 2,560,000
SC4-34 34 60,000 640,000 1,280,000 5,120,000
SC4-37 37 23,4000 250,000 500,000 2,000,000
CP-40 40 307 3,270 6,540 26,160
CP-41 41 1,151 12,280 24,560 98,240
CP-42 42 600 6,400 12,800 51,200
CP-51 51 4,800 51,200 102,400 409,600
CP-52 52 9,216 98,300 196,600 786,400
Spindle Entry Code
LV RV HA HB
SPINDLE RANGE COEFFICIENT
Rheometer Series
Notes: RV spindle ranges are calculated with the RV Guardleg in use.
LV spindle ranges are calculated with the LV Guardleg in use.