Brookfield Engineering Laboratories, Inc. Page 59 Manual No. M/98-211-A0701
Din Adapter
Shear
DAA Rate
Spindle (sec-1) LVDV-III+ RVDV-III+ HADV-III+ HBDV-III+
Viscosity (cP)
85 0 - 258 0.6 - 5,000 6.1 - 5,000 12.2 - 5,000 48.8 - 5,000
86 0 - 258 1.8 - 10,000 18.2 - 10,000 36.5 - 10,000 146 - 10,000
87 0 - 258 5.7 - 50,000 61.0 - 50,000 121 - 50,000 485 - 50,000
Spiral Adapter
Spiral Shear Rate
Spindle (sec-1) LVDV-III+ RVDV-III+ HADV-III+ HBDV-III+
SA-70 .677 - 67.7 98.4 - 98,416 1,050 - 1,050,000 2,100 - 2,100,000 8,400 - 8,400,000
(1 - 100 RPM)
Viscosity (cP)
Cone/Plate Rheometer
Shear
Cone Rate
Spindle (sec-1) LVDV-III+C/P RVDV-III+C/P HADV-II+IC/P HBDV-II+IC/P
Viscosity (cP)
CPE-40 0 - 1500 0.15 - 3,065 1.7 - 32,700 3.3 - 65,400 13.1 - 261,000
CPE-41 0 - 400 0.58 -11,510 6.2 - 122,800 12.3 - 245,600 49.1 - 982,400
CPE-42 0 - 760 0.3 - 6,000 3.2 - 64,000 6.4 - 128,000 25.6 - 512,000
CPE-51 0 - 768 2.4 -47,990 25.6 - 512,000 51.2 -1,024,000 205.0 -4,096,000
CPE-52 0 - 400 4.6 -92,130 49.2 - 983,000 98.3 -1,966,000 393.0 -7,864,000
Helipath with T-Bar Spindles
T-Bar
Spindle LVDV-III+ RVDV-III+ HADV-III+ HBDV-III+
T-A 156 - 187,460 2,000 - 200,000 4,000 - 4,000,000 16,000 - 16,000,000
T-B 312 - 374,920 4,000 - 400,000 8,000 - 8,000,000 32,000 - 32,000,000
T-C 780 - 937,300 10,000 - 10,000,000 20,000 - 20,000,000 80,000 - 80,000,000
T-D 1,560 -1,874,600 20,000 - 20,000,000 40,000 - 40,000,000 160,000 - 160,000,000
T-E 3,900 -4,686,500 50,000 - 50,000,000 100,000 -100,000,000 400,000 - 400,000,000
T-F 7,800 -9,373,000 100,000 -100,000,000 200,000 -200,000,000 800,000 - 800,000,000
Viscosity (cP)
In taking viscosity measurements with the DV-III+ Rheometer there are two considerations which
pertain to the low viscosity limit of effective measurement.
1) Viscosity measurements should be accepted within the equivalent % Torque Range from
10% to 100% for any spindle/speed combination.
2) Viscosity measurements should be taken under laminar flow conditions, not under turbulent
flow conditions.
The first consideration has to do with the precision of the instrument. All DV-III+ Rheometers have a