Vane Spindles
Assembly & Operating Instructions
Manual No. M02-290-C0809
Primary Immersion Mark
Secondary Immersion Mark
Vane
Spindle
Choose the immersion mark which best suits
the size of container for your sample material.
Vane Spindle Set
includes 3 vane spindles
and storage case
Part No. SSVANE
(empty slot)
Vane spindles can be used with any standard Brookeld Viscometer/Rheometer. V-71, V-72, and V-73 are
supplied in the Vane Spindle Set (Part No. SSVANE). The V-74 and V-75 spindles are optional extras. The
small size of the V-74 vane spindle is ideal for testing small sample sizes and tting into sample containers
with small openings. However, limited surface area for spindle measurement geometry may make it more
challenging to obtain repeatable shear stress and viscosity values.
Table 1 - Vane Spindle Dimensions
Spindle
Vane Length
inches cm
Vane Diameter
inches cm
V-71 2.708 6.878 1.354 3.439
V-72 1.708 4.338 .853 2.167
V-73 .998 2.535 .499 1.267
V-74 .463 1.176 .232 .589
V-75 .632 1.61 .316 .803
Note: Sample container diameter should be at least twice (2x) the vane
diameter when possible. Sample container depth should provide
clearance, at the bottom, greater than or equal to vane spindle
diameter when possible.