Chapter 7
Operation
UPK – Define the
active parameter
set
For the typical reader parameters (parameter number 100 to max. 240), up to 8
parameter sets can be created and selected. After starting the reader, the default
parameter set will automatically become the active parameter set. Any changes
to the reader parameters only affect the active parameter set. The active
parameter set can be determined by the UPK command.
Example:
<< n0 (Powerup reset)
>> UGK0
<< ugk000
Active parameter set number 00 (default value)
>> UPK007
<< upk000
New active parameter set was successfully defined
>> UGK0
<< ugk007
Active parameter set number 07
>> P06A84
<< p0
Transmission output of Port 1 changed from 0x6c to 0x84 (parameter set 7!)
...
Poweroff reset
...
<< n0
>> UGK0
<< ugk000
>> F06A
<< f06A6C
The default parameter is loaded after a reset
Host reader
CMD Reader ID Parameter set no.
UPK 1 byte 2 bytes
Reader host
CMD Reader ID Result
upk 1 byte 2 bytes