User's Manual

Table Of Contents
Product Manual - RFID Reader UF70 Certum 160
Chapter 8
Service and Troubleshooting
1) because the device is still busy or because a message has not been confirmed
by the previous read up to now.
8.5 Error display with LED
Power LED does
not light up
8.6 Reader does not respond
B no ackn the message which has
to be confirmed has
been sent the maximum
number of times (rs232
maxrepeat), and has
not been confirmed by
the terminal within the
defined time frame (see
parameter 5)
Host system does not
acknowledge the
message
check availability of host
system (terminal); check
RS232 cables and
connectors
Error
ID
Name Description Possible causes Correcting action
Check the power supply and the connecting cables!
If the LED does not light up, disconnect the device from the power supply
and carefully remove the fuse! Test the fuse. If it is defective, replace it
with a fuse specified by the manufacturer!
If the above measures do not solve the problem, leave the device
disconnected and contact the manufacturer!
Check if the interface connection cable is undamaged and correctly
connected to both reader and host!
Check the status as indicated by the LED!
If you are unsure of the enabled communication interface, try both (RS232
and Ethernet)!
Contact Brooks Automation (Germany) GmbH regarding the firmware of
the device and the firmware update software!