Section 3 Operation
Models MT 3809 and 3819
Installation and Operation Manual
X-VA-MT3809-3819-eng
Part Number: 541B049AHG
September, 2008
F1 F4F3F2
SAVE
Brooks MT 38xxVA
Flow
1 FlowRate Units
2 Configuration
3 Meter Calibration
F1 F4F3F2
ENTER
Brooks MT 38xxVA
FlowRate Units
gal/min
L/min
I mpgal/min
ESC
F1 F4F3F2
SAVE
Brooks MT 38xxVA
On-line
5 LRV
1 Device setup
2 PV
3 AO
4 URV
50.00 gpm
setup
25 gpm
12.0 mA
0.00 gpm
From the Flow Menu, select
FlowRate Units of Measure
(FlowRate UOM).
In Flow, scroll through list
optional units. Highlight the
FlowRate UOM desired. Press
ENTER (F4) to enter the new
information, and press SEND
(F2) to send this information to
the transmitter.
Note: Two Warning messages will appear.
Warning Pressing OK will change device output. Put loop in manual.
Warning Return control loop to automatic control.
These warnings confirm changes to the programming and alert the user to
make changes in other related fields.
Example 2: Programming Low Flow Cutoff.
From the Online Menu, select
Device Setup.