Installation and Operation Manual X-VA-MT3810-eng Part Number: 541B050AAG December, 2008 Model MT 3810 Brooks® Model MT 3810 Metal Tube Variable Area Flowmeter with Optional Electronics Based on Smart Meter ManagerTM Technology Brooks® Model MT 3810
Model MT 3810 Installation and Operation Manual X-VA-MT3810-eng Part Number: 541B050AAG December, 2008 Essential Instructions Read this page before proceeding! Brooks Instrument designs, manufactures and tests its products to meet many national and international standards. Because these instruments are sophisticated technical products, you must properly install, use and maintain them to ensure they continue to operate within their normal specifications.
Installation and Operation Manual X-VA-MT3810-eng Part Number: 541B050AAG December, 2008 Model MT 3810 Dear Customer, We appreciate this opportunity to service your flow measurement and control requirements with a Brooks Instrument device. Every day, flow customers all over the world turn to Brooks Instrument for solutions to their gas and liquid low-flow applications.
Installation and Operation Manual X-VA-MT3810-eng Part Number: 541B050AAG December, 2008 Model MT 3810 THIS PAGE WAS INTENTIONALLY LEFT BLANK
Installation and Operation Manual X-VA-MT3810-eng Part Number: 541B050AAG December, 2008 Contents Model MT 3810 Section 1 Introduction Paragraph Page Number Number 1-1 Description ....................................................................................................................................... 1-1 1-2 Design Features ............................................................................................................................... 1-1 1-3 Specifications - Meter ..........
Contents Model MT 3810 Installation and Operation Manual X-VA-MT3810-eng Part Number: 541B050AAG December, 2008 Figures Figure Page Number Number 1-1 Model MT 3810 Meter Dimensions................................................................................................... 1-4 1-2 Transmitter Only Wiring Diagram ..................................................................................................... 1-7 1-3 Transmitter Only Wiring Notes .................................................
Section 1 Introduction Installation and Operation Manual X-VA-MT3810-eng Part Number: 541B050AAG December, 2008 Model MT 3810 1-1 Description The Brooks® Model MT 3810 Variable Area Flowmeter is a rugged, all metal flowmeter offering 5% full scale accuracy. The Model MT 3810 is constructed with stainless steel components for measuring a variety of liquid and gas applications.
Section 1 Introduction Installation and Operation Manual X-VA-MT3810-eng Part Number: 541B050AAG December, 2008 Model MT 3810 Scales Standard: Detachable aluminum plate single or dual scales Graduations: Choice of direct reading units, millimeter or percentage of maximum flow. Operating Fluid Temperature Limits/Meter only Minimum: -58°F (-50°C) Maximum: 420°F (215°C) Meter with Valve: 392oF (200oC) Refer to Table 1-3 for temperature limitations for meters with electronics.
Section 1 Introduction Installation and Operation Manual X-VA-MT3810-eng Part Number: 541B050AAG December, 2008 Model MT 3810 Table 1-1 Model MT 3810 Capacities, Pressure Drop and Viscosity Immunity Ceiling METER SIZE CONNECTION SIZE DIN ANSI DN mm inches 15 1/2" 7 15 1/2" 25 1" 40 1 1/2" 50 2" 8 10 12 13 FLOAT CODE A B* C D* A B C D A B C D A B C D A B C D WATER l/h gpm 25 0.11 65 0.28 130 0.59 200 0.88 250 1.1 400 1.7 650 2.8 1000 4.4 1200 5.2 1500 6.
Section 1 Introduction Installation and Operation Manual X-VA-MT3810-eng Part Number: 541B050AAG December, 2008 Model MT 3810 Model MT 3810 NPT-F Connection mm (inches)* Connection A B C D E WEIGHT (Approx.) 7&8 1/2" NPT-F 225 (8.85) 88 (3.46) 72 (2.83) 53 (2.09) 59 (2.32) 1.3 kg (3 lbs.) 10 1" NPT-F 300 (11.81) 95 (3.78) 80 (3.15) 59 (2.32) 96 (3.78) 2.8 kg (7 lbs.) 12 1-1/2" NPT-F 300 (11.81) 105 (4.13) 89 (3.50) 65 (2.56) 96 (3.78) 5 kg. (12 lbs.) 13 2" NPT-F 300 (11.
Section 1 Introduction Installation and Operation Manual X-VA-MT3810-eng Part Number: 541B050AAG December, 2008 Model MT 3810 1-5a Microprocessor Transmitter with or without Alarms and Pulse Output Based on Brooks Smart Meter Manager Technology Design Features • A 2-wire, loop-powered device for ease of wiring and installation • 4-20 mA analog output for flowrate, with Bell-202 modulated HART communication channel • User selectable 0% and 100% analog output ranges with optional smoothing • Flexible (mix
Section 1 Introduction Installation and Operation Manual X-VA-MT3810-eng Part Number: 541B050AAG December, 2008 Model MT 3810 Specifications - SMM Microprocessor Transmitter with or without Alarm and Pulse Output EMC Directive 89/336/EEC: EN 50081, EN 50082 and EN 61326-1 Hazardous Location Classification Enclosure: Type 4X/ IP65 Ambient Temperature: -22°F> Tamb < 150°F (-30°C > Tamb < 65°C) Intrinsically Safe United States and Canada UL Listed, E73889, Vol. 1, Sect.
Installation and Operation Manual X-VA-MT3810-eng Part Number: 541B050AAG December, 2008 Section 1 Introduction Model MT 3810 Ref. No.981Y006 - Rev.
Section 1 Introduction Model MT 3810 Installation and Operation Manual X-VA-MT3810-eng Part Number: 541B050AAG December, 2008 Ref. No.981Y006 - Rev.
Installation and Operation Manual X-VA-MT3810-eng Part Number: 541B050AAG December, 2008 Section 1 Introduction Model MT 3810 Ref. No.981Y005 - Rev.
Section 1 Introduction Model MT 3810 Installation and Operation Manual X-VA-MT3810-eng Part Number: 541B050AAG December, 2008 Ref. No.981Y005 - Rev.
Section 1 Introduction Installation and Operation Manual X-VA-MT3810-eng Part Number: 541B050AAG December, 2008 Model MT 3810 Explosion- proof/ Flame-proof United States and Canada UL Listed, E73889, Vol. 1, Sect. 14 Class I, Division 1, Groups C, D; Dust Ignition-proof, Class II, Division 1, Groups E, F, G; Class III; T4 Europe – KEMA 01ATEX2207 X II 2 G EEx d IIB T4 II 2 D T135o C Note: Load resistance should also include I.S. barrier resistance.
Section 1 Introduction Installation and Operation Manual X-VA-MT3810-eng Part Number: 541B050AAG December, 2008 Model MT 3810 Linearity Less than 1% at I maximum Temperature Influence Less than 0.04% per °C Voltage Influence Less than 0.002%/Vdc Load Resistance Influence ± 0.1% full scale Transmitter, Alarm and Pulse Output Wiring Diagrams Refer to Figures 1-2, 1-3, 1-4, 1-5 and 1-7 For Division 1 explosion proof installations, the optional explosion proof enclosure must be used.
Installation and Operation Manual X-VA-MT3810-eng Part Number: 541B050AAG December, 2008 Section 1 Introduction Model MT 3810 Ref. No.981Y008 - Rev.
Section 1 Introduction Installation and Operation Manual X-VA-MT3810-eng Part Number: 541B050AAG December, 2008 Model MT 3810 1-5b Microprocessor Transmitter with Inductive Alarms This combined system provides both the sophistication of the microprocessor along with the simplicity of one or two switch inductive alarms. Specifications for the transmitter are as stated previously and specifications for the front adjustable inductive alarms are as follows.
Installation and Operation Manual X-VA-MT3810-eng Part Number: 541B050AAG December, 2008 Section 1 Introduction Model MT 3810 Ref. No.981Y007 - Rev.
Section 1 Introduction Model MT 3810 Installation and Operation Manual X-VA-MT3810-eng Part Number: 541B050AAG December, 2008 Ref. No.981Y007 - Rev.
Section 1 Introduction Installation and Operation Manual X-VA-MT3810-eng Part Number: 541B050AAG December, 2008 Model MT 3810 1-5c Inductive Alarm Switches Design Features • 1 or 2 normally open inductive limit switches • Optional intrinsically safe power supply/amplifier/relay unit • For low or high limit signaling/switching • Front adjustable Description 1 or 2 electronic limit switches can be installed in the indicator housing to allow initiation of signaling or switching functions on a preset flow va
Section 1 Introduction Installation and Operation Manual X-VA-MT3810-eng Part Number: 541B050AAG December, 2008 Model MT 3810 Explosion- proof/ Flame-proof United States and Canada UL Listed, E73889, Vol. 1, Sect. 14 Class I, Division 1, Groups C, D; Dust Ignition-proof, Class II, Division 1, Groups E, F, G; Class III; T4 Europe – KEMA 01ATEX2207 X II 2 G EEx d IIB T4 II 2 D T135oC Power Supply 5-25 Vdc; 25 mA max.
Installation and Operation Manual X-VA-MT3810-eng Part Number: 541B050AAG December, 2008 Section 1 Introduction Model MT 3810 Ref. No.981Y001 - Rev.
Section 1 Introduction Model MT 3810 Installation and Operation Manual X-VA-MT3810-eng Part Number: 541B050AAG December, 2008 Ref. No.981Y001 - Rev.
Section 2 Installation Installation and Operation Manual X-VA-MT3810-eng Part Number: 541B050AHG June, 2008 Model MT 3810 2-1 General This section contains the procedures for the receipt and installation of the instrument. See Section 1 for dimensional and connection requirements. Do not attempt to start the system until the instrument has been permanently installed. It is important that the start-up procedures be followed in the exact sequence presented.
Section 2 Installation Installation and Operation Manual X-VA-MT3810-eng Part Number: 541B050AHG June, 2008 Model MT 3810 2-4 Return Shipment Prior to returning any instrument to the factory, contact your nearest Brooks location for a Return Materials Authorization Number (RMA#). This can be obtained from one of the following locations: Brooks Instrument 407 W. Vine Street P.O.
Section 2 Installation Installation and Operation Manual X-VA-MT3810-eng Part Number: 541B050AHG June, 2008 Model MT 3810 2-7 Installation of Flowmeter WARNING If the inlet and outlet valves adjacent to the indicator are to be closed for any reason, the indicator must be completely drained. Failure to do so may result in thermal expansion of the liquid which can rupture the meter and cause possible personal injury. Recommended installation for Models MT 3809 and MT 3819 is as follows: a.
Section 2 Installation Model MT 3810 Installation and Operation Manual X-VA-MT3810-eng Part Number: 541B050AHG June, 2008 2-8 Installation of Model MT 3810 Flowmeter with a Smart Meter Manager Transmitter with or without Optional Alarms and Pulse Output a. Install the meter as described in Section 2-4. b. Transmitter setup is done solely through proper system wiring and parameter configuration. Common transmitter parameters are set by Brooks prior to shipment.
Section 2 Installation Installation and Operation Manual X-VA-MT3810-eng Part Number: 541B050AHG June, 2008 Model MT 3810 ® inside TM Smart Meter Manager Figure 2-2 Smart Meter Manager Electrical Configuration ® inside TM Smart Meter Manager Figure 2-3 Typical SMM Transmitter Analog Output and Power Wiring 2-5
Section 2 Installation Installation and Operation Manual X-VA-MT3810-eng Part Number: 541B050AHG June, 2008 Model MT 3810 ® inside TM Smart Meter Manager Figure 2-4 Alternate SMM Transmitter Analog Output and Power Wiring (Where 4-20 mA signal is not required) ® inside TM Smart Meter Manager ® inside TM Smart Meter Manager Figure 2-5 Multi-drop SMM Transmitter Analog Output and Power Wiring 2-6
Installation and Operation Manual X-VA-MT3810-eng Part Number: 541B050AHG June, 2008 Section 2 Installation Model MT 3810 Typically applications require only the use of the 2 wire loop analog signal. In some applications where transmitters and actuators are widely separated (e.g. tank farms), devices are wired in a multi-drop configuration to save wiring costs. Each unit is given an individual HART address in the range of 1-15 to distinguish each unit during communications over a common wire pair.
Section 2 Installation Installation and Operation Manual X-VA-MT3810-eng Part Number: 541B050AHG June, 2008 Model MT 3810 7. Under actual flow conditions, verify that the transmitter output matches the mechanical pointer position. If a discrepancy is noted, the HART communications channel can be used to verify or adjust the transmitter settings. The Emerson Model 275 hand held communicator provides a simple means to accomplish this check anywhere in the loop.
Installation and Operation Manual X-VA-MT3810-eng Part Number: 541B050AHG June, 2008 Section 2 Installation Model MT 3810 All internal parameters can be accessed over the HART communications channel, including the configuration of the output signals used for alarms and pulse output. An external termination box is attached for easy hookup of these signals. The alarms are configurable as normally open or normally closed.
Section 2 Installation Model MT 3810 Installation and Operation Manual X-VA-MT3810-eng Part Number: 541B050AHG June, 2008 8. After installation and powering of the loop, the transmitter must be zeroed, both electrically and mechanically. This operation will compensate for any stray magnetic effects in the vicinity of the transmitter. Flow must be zero and when the zero function is momentarily activated.
Installation and Operation Manual X-VA-MT3810-eng Part Number: 541B050AHG June, 2008 Section 2 Installation Model MT 3810 To install the inductive alarms (1 or 2 switches): 1. The inductive alarms can be supplied as a stand alone option or combined with the transmitter. When stand alone, connections to the alarms are made inside the indicator housing. When supplied in combination with the transmitter, connections are made inside the auxiliary terminal box attached to the indicator housing.
Section 2 Installation Model MT 3810 Installation and Operation Manual X-VA-MT3810-eng Part Number: 541B050AHG June, 2008 8. If desired, the transmitter zero function can be remotely mounted using a momentary contact switch. Follow the same wiring practices as used for the other outputs. 9. After installation and powering of the loop, the transmitter must be zeroed, both electrically and mechanically. This operation will compensate for any stray magnetic effects in the vicinity of the transmitter.
Section 3 Operation Installation and Operation Manual X-VA-MT3810-eng Part Number: 541B050AHG December, 2008 Model MT 3810 3-1 Pre-Start Check After the flowmeter has been properly installed in the process, it is ready for operation. When initiating flow, slowly open the valve to avoid a flow surge.Bypass is a help in bringing the flow on smoothly.
Section 3 Operation Installation and Operation Manual X-VA-MT3810-eng Part Number: 541B050AAG December, 2008 Model MT 3810 Transmitter Only (uP) Pre-Programmed Parameters Transmitter (uP) Plus...
Section 3 Operation Installation and Operation Manual X-VA-MT3810-eng Part Number: 541B050AHG December, 2008 Model MT 3810 Device Description Process Variables PV (FlowRate) AO Percent SV (ResettableTotalizer) TV (InventoryTotalizer) QV (unused?) Diagnostics & Service Alarm Status Diagnostic Code Sequence Number Current Float Position Internal Temperature Test Device Analog/Loop Test Set 4mA Set 20mA End Test Contact Out1 Test Set ON Set OFF End Test Contact Out2 Test Set ON Set OFF End Test Pulse Outp
Section 3 Operation Installation and Operation Manual X-VA-MT3810-eng Part Number: 541B050AAG December, 2008 Model MT 3810 Detailed Setup Measurements Computations Totalizers Flow FlowRate UOM Meter Configuration Meter Zero LowFlow Cutoff Calibration Factor Meter Calibration CalibFlowRate UOM Calibration Points Float Position#1 FlowRate#1 : Float Position#10 FlowRate#10 Calibration Info Process Fluid Calibration Tool Calibrator Name Calibration Date Calib Temperature UOM Calibration Temperature Gas Pi
Section 3 Operation Installation and Operation Manual X-VA-MT3810-eng Part Number: 541B050AHG December, 2008 Model MT 3810 The HART Communicator is the hand-held interface that provides a common communication link to all HART compatible, microprocessorbased devices. A keypad, liquid crystal display (LCD) and software menu structure make up the HART communicator user interface. The Emerson Model 275 is easy to use. Just press a few buttons to become familiar with the Action Keys and menu structure.
Section 3 Operation Installation and Operation Manual X-VA-MT3810-eng Part Number: 541B050AAG December, 2008 Model MT 3810 Emerson Model 275 Communicator Functions Action Keys and Hot Key Six Action Keys promote easy navigation through the menu structure. You may customize the Hot Key to quickly access a menu of your most frequently performed on-line tasks. The Hot Key Menu is a user-definable menu containing one permanent option, Range Values. Range Values provides quick access to rearrange capability.
Section 3 Operation Installation and Operation Manual X-VA-MT3810-eng Part Number: 541B050AHG December, 2008 Model MT 3810 Function Keys Use the four function keys, marked F1 through F4 located below the LCD to perform software functions as indicated by the dynamic labels. Different labels appear over the four function keys as you move among the various menus. On-line Menu When connected to a HART compatible device, simply press the ON Key to display the On-line Menu.
Section 3 Operation Model MT 3810 Installation and Operation Manual X-VA-MT3810-eng Part Number: 541B050AAG December, 2008 Computations: Totalization Totalizers are available in both Resettable and Inventory options. Resettable totalization is used for batching while inventory totalization measures the total volume over time. Units of measure are set independently for each of these options.
Section 3 Operation Installation and Operation Manual X-VA-MT3810-eng Part Number: 541B050AHG December, 2008 Model MT 3810 The following are two examples of typical programming via a Emerson Model 275 hand-held communicator. Example 1: Programming Flow Rate Units. Brooks MT 38xxVA On-line 1 Device Setup setup 2 PV 25 gpm 3 AO 12.0 mA 4 URV 0.00 gpm 5 LRV 50.00 gpm SAVE F1 F2 F3 From the Online Menu, select Device Setup.
Section 3 Operation Installation and Operation Manual X-VA-MT3810-eng Part Number: 541B050AAG December, 2008 Model MT 3810 Brooks MT 38xxVA Measurements 1 Flow 2 Temperature 3 Density 4 Pressure 5 Viscosity SAVE F1 F2 From the Measurements Menu, select Flow. F3 F4 Brooks MT 38xxVA Flow 1 FlowRate Units 2 Configuration 3 Meter Calibration From the Flow Menu, select FlowRate Units of Measure (FlowRate UOM).
Section 3 Operation Installation and Operation Manual X-VA-MT3810-eng Part Number: 541B050AHG December, 2008 Model MT 3810 Example 2: Programming Low Flow Cutoff. Brooks MT 38xxVA On-line 1 Device setup setup 2 PV 25 gpm 3 AO 12.0 mA 4 URV 0.00 gpm 5 LRV 50.00 gpm SAVE F1 F2 F3 F4 Brooks MT 38xxVA Device Setup 1 Process Variables 2 Diags & Services 3 Basic Setup 4 Detailed Setup From the Device Setup Menu, select Basic Setup.
Section 3 Operation Installation and Operation Manual X-VA-MT3810-eng Part Number: 541B050AAG December, 2008 Model MT 3810 Brooks MT 38xxVA Configuration 1 Cutoff 2 CalFact From the Configuration Menu, select Cutoff. SAVE F1 F2 F3 F4 Brooks MT 38xxVA Cutoff 00.00 L/min HELP F1 3-12 DEL F2 ESC F3 ENTER F4 Enter the Cutoff value in the same units specified in the FlowRate UOM. Press ENTER (F4) to enter the new information and press SEND (F2) to send this information to the transmitter.
Section 3 Operation Installation and Operation Manual X-VA-MT3810-eng Part Number: 541B050AHG December, 2008 Model MT 3810 3-4 Operation of the Model MT 3810 Flowmeter with a Smart Meter Manager Transmitter with Inductive Alarms ( 1 or 2 Switches) a. Start-up the meter as described in Section 3-2. b. Start-up the transmitter as described in Section 3-3. c. Start-up the inductive alarms as described in Section 3-5. 3-5 Operation of the Model MT 3810 Flowmeter with Inductive Alarms ( 1 or 2 Switches) a.
Section 3 Operation Installation and Operation Manual X-VA-MT3810-eng Part Number: 541B050AAG December, 2008 Model MT 3810 THIS PAGE WAS INTENTIONALLY LEFT BLANK 3-14
Installation and Operation Manual X-VA-MT3810-eng Part Number: 541B050AAG December, 2008 Section 4 Maintenance Model MT 3810 4-1 General Service Information 4-1
Section 4 Maintenance Installation and Operation Manual X-VA-MT3810-eng Part Number: 541B050AAG December, 2008 Model MT 3810 There is no routine maintenance required for the Models MT 3810 flowmeters. However should it become necessary to replace, adjust or remove components of the flowmeter, specific recommendations must be followed. 4-2 Meter Float Replacement and Cleaning Float replacement procedures are dependent on the meter size, connection type and fluid (gas vs. liquid) application.
Installation and Operation Manual X-VA-MT3810-eng Part Number: 541B050AAG December, 2008 Section 4 Maintenance Model MT 3810 5. Reassemble by inserting the float assembly into the top of the meter. Replace the guide vane (#3) and secure the float assembly by tightening the inlet and outlet screws (#2). Replace the spiral retaining ring (#1) in the meter inlet. c. Size 8 (1/2") Flanged Gas Service 1. Remove the meter from the process line and lay the meter horizontal on a table. 2.
Section 4 Maintenance Model MT 3810 Installation and Operation Manual X-VA-MT3810-eng Part Number: 541B050AAG December, 2008 f. Sizes 15 and 16 (3" and 4") Flanged Liquid or Gas Service (Size 15 only) 1. Remove the meter from the process line and lay the meter horizontal on a table. 2. Loosen the float assembly by fixating two pliers one on each end of the float assembly inlet guide cartridge (#8). Turn to loosen with the help of a pin wrench. 3.
Section 4 Maintenance Installation and Operation Manual X-VA-MT3810-eng Part Number: 541B050AAG December, 2008 Model MT 3810 d. Sizes 10 (1"), 12 (1 1/2"), 13 (2") NPT Gas Service 1. Remove the meter from the process line and lay the meter horizontal on a table. 2. Remove the top/outlet screw (#5). 3. Loosen the float assembly by holding the damper in place with a pin (1/8” diameter) in the top while loosening the inlet/bottom fitting. Gently push the float assembly out through the inlet/bottom. 4.
Section 4 Maintenance Installation and Operation Manual X-VA-MT3810-eng Part Number: 541B050AAG December, 2008 Model MT 3810 Liquid Service Gas Service Figure 4-1 Meter Float Replacement & Cleaning 4-6
Installation and Operation Manual X-VA-MT3810-eng Part Number: 541C050AAG December, 2008 Appendix A CE Certification Brooks® Model MT 3810 Dansk Brooks Instrument 407 West Vine St. Hatfield, PA 19440 U.S.A. Emne : Tillæg til instruktions manual. Reference : CE mærkning af Masse Flow udstyr Dato : Januar-1996. Brooks Instrument har gennemført CE mærkning af elektronisk udstyr med succes, i henhold til regulativet om elektrisk støj (EMC direktivet 89/336/EEC).
Appendix A CE Certification Brooks® Model MT 3810 Installation and Operation Manual X-VA-MT3810-eng Part Number: 541C050AAG December, 2008 English Brooks Instrument 407 West Vine St. Hatfield, PA 19440 U.S.A. Subject : Addendum to the Instruction Manual. Reference : CE certification of Mass Flow Equipment Date : January-1996. The Brooks (electric/electronic) equipment bearing the CE mark has been successfully tested to the regulations of the Electro Magnetic Compatibility (EMC directive 89/336/EEC).
Installation and Operation Manual X-VA-MT3810-eng Part Number: 541C050AAG December, 2008 Appendix A CE Certification Brooks® Model MT 3810 Français Brooks Instrument 407 West Vine St. Hatfield, PA 19440 U.S.A. Sujet : Annexe au Manuel d’Instructions. Référence : Certification CE des Débitmètres Massiques à Effet Thermique. Date : Janvier 1996.
Appendix A CE Certification Brooks® Model MT 3810 Installation and Operation Manual X-VA-MT3810-eng Part Number: 541C050AAG December, 2008 Italiano Brooks Instrument 407 West Vine St. Hatfield, PA 19440 U.S.A. Oggetto : Addendum al manuale di istruzioni. Riferimento : Certificazione CE dei misuratori termici di portata in massa Data : Gennaio 1996.
Installation and Operation Manual X-VA-MT3810-eng Part Number: 541C050AAG December, 2008 Appendix A CE Certification Brooks® Model MT 3810 Norsk Brooks Instrument 407 West Vine St. Hatfield, PA 19440 U.S.A.
Appendix A CE Certification Brooks® Model MT 3810 Installation and Operation Manual X-VA-MT3810-eng Part Number: 541C050AAG December, 2008 Suomi Brooks Instrument 407 West Vine St. Hatfield, PA 19440 U.S.A. Asia : Lisäys Käyttöohjeisiin Viite : Massamäärämittareiden CE sertifiointi Päivämäärä : Tammikuu 1996 Brooksin CE merkillä varustetut sähköiset laitteet ovat läpäissyt EMC testit (direktiivi 89/336/EEC). Erityistä huomiota on kuitenkin kiinnitettävä signaalikaapelin valintaan.
Installation and Operation Manual X-VA-MT3810-eng Part Number: 541B050AAG December, 2008 Models MT 3810 THIS PAGE WAS INTENTIONALLY LEFT BLANK
Installation and Operation Manual X-VA-MT3810-eng Part Number: 541B050AAG December, 2008 Models MT 3810 LIMITED WARRANTY Seller warrants that the Goods manufactured by Seller will be free from defects in materials or workmanship under normal use and service and that the Software will execute the programming instructions provided by Seller until the expiration of the earlier of twelve (12) months from the date of initial installation or eighteen (18) months from the date of shipment by Seller.