Installation and Operation Manual X-PR-SolidSense II-PT-eng Part Number: 541B160AAG July, 2011 SolidSense II® Series Brooks® SolidSense II® Series Pressure Transducer SolidSense II Pressure Transducer 9-Pin Flow Thru Configuration SolidSense II Pressure Transducer Pigtail Configuration SolidSense II Pressure Transducer with Bendix® Configuration SolidSense II Pressure Transducer with Readout Configuration
Installation and Operation Manual X-PR-SolidSense II-PT-eng Part Number: 541B160AAG July, 2011 SolidSense II® Series Essential Instructions Read before proceeding! Brooks Instrument designs, manufactures and tests its products to meet many national and international standards. These products must be properly installed, operated and maintained to ensure they continue to operate within their normal specifications.
Installation and Operation Manual X-PR-SolidSense II-PT-eng Part Number: 541B160AAG July, 2011 SolidSense II® Series Dear Customer, We appreciate this opportunity to service your pressure measurement and control requirements with a Brooks Instrument device. Every day, customers all over the world turn to Brooks Instrument for solutions to their gas and liquid control applications.
Installation and Operation Manual X-PR-SolidSense II-PT-eng Part Number: 541B160AAG July, 2011 SolidSense II® Series THIS PAGE WAS INTENTIONALLY LEFT BLANK
Installation and Operation Manual X-PR-SolidSense II-PT-eng Part Number: 541B160AAG July, 2011 Paragraph Number Contents SolidSense II® Series Page Number Section 1 Introduction 1-1 Introduction ................................................................................................................................... 1-1 1-2 Intended Use .................................................................................................................................
Contents SolidSense II® Series Installation and Operation Manual X-PR-SolidSense II-PT-eng Part Number: 541B160AAG July, 2011 Figures Figure Number 1-1 1-2A 1-2B 1-2C 1-2D 1-3 2-1 2-2 2-3 2-4 2-5 2-6 2-7 2-8 2-9 2-10 ii Page Number SolidSense II Pressure Transducer Configurations ...................................................................... 1-1 SolidSense II Pressure Transducer Dimensions for Thru-Tubes with DB15, Bendix and Pigtail Connections ................................................
Installation and Operation Manual X-PR-SolidSense II-PT-eng Part Number: 541B160AAG July, 2011 Section 1 Introduction SolidSense II® Series 1-1 Introduction This manual covers the Brooks® SolidSense II® Series pressure transducers as shown in Figure 1-1 below. Included herein is general information, operating specifications, installation, removal instructions and product warranty information.
Section 1 Introduction SolidSense II® Series Installation and Operation Manual X-PR-SolidSense II-PT-eng Part Number: 541B1560AAG July, 2011 1-2 Intended Use The Brooks SolidSense II Series pressure transducers meet the most rigid semiconductor and industrial high purity application requirements. With accuracy of 0.25% of full scale, SolidSense II pressure transducers are used to provide years of reliable pressure monitoring in a variety of applications.
Section 1 Introduction Installation and Operation Manual X-PR-SolidSense II-PT-eng Part Number: 541B160AAG July, 2011 SolidSense II® Series 1-4 Product Description Code The following Product Description Code (PDC) identifies the SolidSense II pressure transducer in various configurations. This code is used to order your SolidSense II pressure transducer. Call your Brooks Service and Support Representative if you need assistance. Code Description I.
Section 1 Introduction Installation and Operation Manual X-PR-SolidSense II-PT-eng Part Number: 541B1560AAG July, 2011 SolidSense II® Series 1-5 SolidSense II Pressure Transducer Specifications Category Electrical Supply Current Power Requirement Operating Guidelines Max. 10 mA for 0.00 Vdc and 0.05 to 5.05 Vdc Output 10 to 30 Vdc for 4 to 20 mA Output 11 to 30 Vdc for 0.05 to 5.05 Output 13 to 32 Vdc for 0.00 to 10.00 Output Connections 4-Pin Bendix connector 6 ft.
Section 1 Introduction Installation and Operation Manual X-PR-SolidSense II-PT-eng Part Number: 541B160AAG July, 2011 SolidSense II® Series DB15 BENDIX PIGTAIL PIGTAIL .84 [21.33] .84 [21.33] .84 [21.33] .84 [21.33] 2.78 [70.64] 3.36 [85.40] 3.36 [85.40] 3.28 [83.23] 2.60 [66.04] 2.96 [75.17] 2.24 [56.90] VCR FIXED MALE THRUTUBE BENDIX GFFXXXBVM FITTING A333125004 VCR SWIVEL FEMALE THRUTUBE DB15 GFFXXXDSF .84 [21.33] BENDIX 4.035[102.
Section 1 Introduction Installation and Operation Manual X-PR-SolidSense II-PT-eng Part Number: 541B1560AAG July, 2011 SolidSense II® Series PIGTAIL PIGTAIL PIGTAIL BENDIX .84 [21.33] 1.06 [26.92] 3.60 [91.54] .84 [21.33] .84 [21.33] 3.36 [85.40] 2.82 [71.62] 3.36 [85.40] 2.94 [74.62] 2X .37 [9.52] .37 [9.52] 2X .31 [7.92] .31 [7.92] .31 [7.92] .25 [6.35] .42 [10.79] 2X .42 [10.79] 1.85 [46.98] 2X .42 [10.
Section 1 Introduction Installation and Operation Manual X-PR-SolidSense II-PT-eng Part Number: 541B160AAG July, 2011 SolidSense II® Series PIGTAIL BENDIX PIGTAIL BENDIX .84 [21.33] .84 [21.33] .84 [21.33] .84 [21.33] 3.74 [95.10] 3.40 [86.36] NO. 4 1.53 [38.97] PIGTAIL .84 [21.33] 1.53 [38.97] W-SEAL BENDIX GFDXXXXBWS GFDXXXXB5W FITTING A333366001 VCR SWIVEL FEMALE PIGTAIL GFDXXXXPSF FITTING A333124002 ME131190 1.48 [37.59] W-SEAL BENDIX GFDXXXXPWS GFDXXXXP5W FITTING A333366001 1.
Section 1 Introduction Installation and Operation Manual X-PR-SolidSense II-PT-eng Part Number: 541B1560AAG July, 2011 SolidSense II® Series PIGTAIL 1.06 [26.92] BENDIX .84 [21.33] 1.06 [26.92] BENDIX 3.42 [86.90] 2.95 [74.98] 3.48 [88.39] 1.12 [28.57] 1 1/8" C-SEAL PIGTAIL 9" TO 15 PIN VCR SWIVEL MALE BENDIX VCR SWIVEL FEMALE BENDIX GFDXXXXBSMA GFDXXXXBSFA GFDXXX3ECS FITTING A333124001 FITTING A333124002 FITTING A333127001 ME131193 ME131190 CABLE 215-31423-000 .84 [21.
Section 1 Introduction Installation and Operation Manual X-PR-SolidSense II-PT-eng Part Number: 541B160AAG July, 2011 SolidSense II® Series PIGTAIL 1.06 [26.92] PIGTAIL PIGTAIL 1.06 [26.92] 1.06 [26.92] 3.60 [91.54] 2.94 [74.62] 4.26 [108.10] 3.67 [93.11] 3.00 [76.20] 3.59 [91.19] .31 [7.92] .25 [6.35] .42 [10.79] 1.85 [46.
Section 1 Introduction Installation and Operation Manual X-PR-SolidSense II-PT-eng Part Number: 541B1560AAG July, 2011 SolidSense II® Series THIS PAGE WAS INTENTIONALLY LEFT BLANK 1-10
Section 2 Installation Installation and Operation Manual X-PR-SolidSense II-PT-eng Part Number 541B160AAG July, 2011 SolidSense II® Series 2-1 General This section provides installation instructions for the Brooks® SolidSense II pressure transducers . Refer to Section 1 of this manual for dimensions. 2-2 Receipt of Equipment When the instrument is received, the outside packing case should be checked for damage incurred during shipment.
Section 2 Installation SolidSense II® Series Installation and Operation Manual X-SolidSense II-PT-eng Part Number 541B160AAG July, 2011 2-4 Return Shipment Prior to returning any device to the factory, visit the Brooks web site (www.BrooksInstrument.com) for a Return Materials Authorization Number (RMA#), or contact one of the locations provided on p. 2-1.
Installation and Operation Manual X-PR-SolidSense II-PT-eng Part Number 541B160AAG July, 2011 Section 2 Installation SolidSense II® Series 2-7 Mechanical Connections Failure to follow these procedures may adversely affect the product’s performance and could void the product warranty. Inspect, but DO NOT unwrap any parts until installation. Contact your Brooks representative with any problems. 1.
Section 2 Installation SolidSense II® Series Installation and Operation Manual X-SolidSense II-PT-eng Part Number 541B160AAG July, 2011 b. Turn the gas flow ON then OFF, 10 times to remove any particles generated during installation. (The flow rate used should at least equal the process flow specifications.) c. Mechanical Installation is complete. Complete the electrical wiring connec-tions as noted in the next section.
Section 2 Installation Installation and Operation Manual X-PR-SolidSense II-PT-eng Part Number 541B160AAG July, 2011 SolidSense II® Series Pin 2 Output Signal Pin 1 Pin 5 Power Supply Common Pin 15 Pin 7 Power Supply Input Pin 12 Output Common Figure 2-3 15-Pin D-Sub HD Connector Wiring Pinout The Bendix connector cable has a removable, 4-pin locking connector with a shielded cable. The connector cable is available from Brooks under part number: EL3000032010.
Section 2 Installation Installation and Operation Manual X-SolidSense II-PT-eng Part Number 541B160AAG July, 2011 SolidSense II® Series Green (+) Output Red (+) Excitation Black (-) Common (Bare) Drain Wire Top key on mating side Red (+) Excitation Female Bendix Connector Black (-) Common A B D C No Connection Green (+) Output (Bare) Drain Wire Solder side view from inside connector Figure 2-5 0-5 Vdc or 0-10 Vdc Bendix Cable Wiring The SolidSense II pressure transducer can be wired directly to a 12
Section 2 Installation Installation and Operation Manual X-PR-SolidSense II-PT-eng Part Number 541B160AAG July, 2011 SolidSense II® Series 2-9 Installation Instructions - Non-Incendive Field Wiring 2-9-1 Non-Incendive Field Wiring 4-20 mA . Notes: 1. The non-incendive field wiring curcuit concept allows interconnection of non-incendive wiring apparatus with associated non-incendive field wiring apparatus using any of the wiring methods permitted for unclassified locations. 2.
Section 2 Installation Installation and Operation Manual X-SolidSense II-PT-eng Part Number 541B160AAG July, 2011 SolidSense II® Series 2-9-2 Non-Incendive Field Wiring 0-5 Vdc . Notes: 1. The non-incendive field wiring curcuit concept allows interconnection of non-incendive wiring apparatus with associated non-incendive field wiring apparatus using any of the wiring methods permitted for unclassified locations. 2. Vmax > Voc or Vt, Ca > Ci + C cable, La > Li + L cable. 3.
Section 2 Installation Installation and Operation Manual X-PR-SolidSense II-PT-eng Part Number 541B160AAG July, 2011 SolidSense II® Series 2-10 Zero Adjustment The SolidSense II pressure transducer is factory calibrated and does not normally need a field adjustment. If a field adjustment becomes necessary, adjust the zero as noted below.
Section 2 Installation Installation and Operation Manual X-SolidSense II-PT-eng Part Number 541B160AAG July, 2011 SolidSense II® Series 2-10-2 Zero Adjustment (ATEX SolidSense II Pressure Transmitters Only) On ATEX SolidSense II pressure transmitters, the zero adjustment screw is located underneath the access cover nut as shown in Figure 2-10(A-B). Unscrew the access cover by turning it in a counterclockwise direction with the electrical connector pointing upward.
Installation and Operation Manual X-PR-SolidSense II-PT-eng Part Number: 541B160AAG July, 2011 SolidSense II Series THIS PAGE WAS INTENTIONALLY LEFT BLANK
SolidSense II Series Installation and Operation Manual X-PR-SolidSense II-PT-eng Part Number: 541B160AAG July, 2011 LIMITED WARRANTY Seller warrants that the Goods manufactured by Seller will be free from defects in materials or workmanship under normal use and service and that the Software will execute the programming instructions provided by Seller until the expiration of the earlier of twelve (12) months from the date of initial installation or eighteen (18) months from the date of shipment by Seller.