Section - VIII Sampling Station BT3000 PLUS Rev. 0, Soft Ver. 8 Page 7 of 10
8-3.3. LEFT SERUM PRIMARY TUBE SENSOR
BOARD (15-04280-01)
This supplementary circuit is located on the left of the serum chamber. It
accommodates a noninvasive infrared detector housed in a PVC body, for the
detection of the primary tube. This circuit board is connected to the Serum Plate
Controller Board (15-04422-01) through the J1 connector, passing through the
another supplementary circuit board on the right. Close by, there is a Hall sensor
module connected to the Serum Hall Sensor Board (15-04282-01) on the serum
plate, which is connected by the J2 connector, to the Serum Primary Tube Sensor
Board (15-04280-01).
The Hall sensor uses the Smith trigger, while the primary tube sensor uses the
U2A-B amplifier in the A.C mode. The emitter LED EM1 is impulse-driven through
the U1C oscillator.
8-3.4. RIGHT SERUM PRIMARY TUBE SENSOR
BOARD (15-04278-01)
This supplementary circuit is located on the right of the serum chamber. This
circuit is identical to the one outlined in the preceding paragraph, but without the
control for the Hall sensor. For operating characteristics, refer to the preceding
paragraph. The J2 connector attaches to the identical board on the left. The J1
connector interconnects these two identical boards through a flat cable to the
Serum Plate Controller Board (15-04422-01).
SERUM MOTOR
I
2
CBUS
SERUM BAR-CODE
VACUTAINER SERUM "R"
SERUM PLATE CONTROLLER MODULE
Figure 8-6
I
2
C BUS
BOARD 15-04422-01