Service manual
Section - VIII Sampling Station BT3000 PLUS Rev. 0, Soft Ver. 8 Page 8 of 10
8-4. BARCODE SCANNER
This is a high quality laser scanning & CCD module (Figures 8-5, & 8-7). It is
connected to the "Serum Plate Controller Module" through the serial port. It can
automatically read 6 different types of codes. It is also possible to reprogram the
scanning modules with additional codes during installation. Bt 3000 PLUS has
barcode scanning modules for the reagent module and the serum module.
Important
Notice:
Any modification to the Variable Serial Protocol (refer to paragraph 4-6. "Variable
Serial Protocol" in Section IV) is restricted to qualified personnel only. The
Biotecnica Instruments S.p.A. guaranties the correct performance of the internal
serial protocol. The responsibility for any malfunction arising out of any
modifications to the scripts of the Variable Serial Protocol rests with the customer.
WARNING
This information regards the setting up of the barcode for sample tubes
identification. The reading of the sample barcode label has the same progression
as patient code.
For example: Once a patient code of 15 characters has been entered, then a code
of 8 characters followed by 7 empty spaces to reach the 15 characters is sent.
The code read on the barcode label must have the same sequence 8 + 7 for
correct detection.
8-5. SERUM & REAGENT BARCODES
The available codes for sampling procedure are in the memory in blocks of 5
codes. To insert a new block (5 codes) it is important to import data using a floppy
disk. The following blocks are available at the present:
CODE39, CODE39, FULL ASCHII, CODABAR, CODE128
EAN128, CODE93, PLESSEY, PHARMACODE, EAN13
UPCA, UPCE, EAN13 AddON2, EAN8 AddOn2, UPCA AddOn2
UPCE AddOn2, EAN13 AddOn5, EAN8 AddOn5, UPCA AddOn5, UPCE AddOn5