Specifications
BT-610 Manual Page 17 BT-610-9800 Rev A
SET CLOCK Set the date and time. Set Clock Screen
SET CONTRAST Adjust the display contrast. Set Contrast Screen
ABOUT Display firmware version and serial number. About Screen
5.1. Edit Main Menu Items
To change settings, press Menu to display the Main Menu, press ▲ or ▼ to navigate to
the desired item and press Enter to display the item view/edit screen.
To edit pick list items (e.g. Sample Setup - Auto/Manual), press ▲ or ▼ to navigate to
the item. Press Enter to select the item. Press ▲ or ▼ to change the setting. Press
ENTER to save the setting or ESC to cancel and return to the main value.
To edit numeric values (e.g. Count Alarms - Alarm Limit), press ▲ or ▼ to navigate to
the item. Press Enter to select the item. Press ▲ or ▼ to increment or decrement a
value. Press ◄ or ► to select the next digit. Press ENTER to save the value or ESC
to cancel and return to the main value.
5.2. Sample Setup Screen
Figure 14 shows the Sample Setup Screen. The 4 parameters are covered in the
following sections.
LOCATION: 001
SAMPLE TIME: 060
HOLD TIME: 000
SAMPLES: SINGLE
Figure 14 – Sample Setup Screen
5.2.1. Location Number
The location number is used to assign a unique number to a location or area. This
important field is included in sample data records (display, printer and serial output).
5.2.2. Sample Time
The sample time determines the amount of time that counts are accumulated while the
pump is running. The length of the sample is user settable from 1 – 999 seconds.
5.2.3. Hold Time
The Hold time is the time between samples when sampling in Auto mode (continuous).
The Hold time is user settable from 0 – 999 seconds. The pump will remain on during
the hold period if the Hold time is 60 seconds or less. The pump will stop after each
sample, and start a few seconds before the next sample, if the Hold time is greater than
60 seconds. Hold times greater than 60 seconds will increase pump life.