To hibernate a test, highlight the test on the left hand side of the screen and click on Put into
Hibernation. To bring a protocol back into the normal list of tests to slect, highlight the test from the right
hand side of the screen and click on Awaken FromHibernation.
Remove Current Case
You may delete client cases by selecting the client/case that you would like to delete, and then clicking
Remove Current Case. This information cannot be retrieved. A warning will appear prior to deleting
data.
Calibrating the Equipment
Before you will be able to use your system you must calibrate and verify the equipment.
If the following message appears when ODES is launched, the pinch grip, hand grip, FOCUS System,
and Pressure Algometer must be calibrated or verified before testing. The message will alert you as to
which tools need to be calibrated.
It is important to note that calibration should be performed on a weekly basis and
verified daily . Prior to an assessment that may be challenged legally, both
calibration and verification of all devices being utilized for the assessment should be
performed.
Prior to calibrating any device, please ensure that the Data Acquisition Box has been set
up successfully and is communicating with your computer. You can check to see if this
is the case by going to Administration | Environment Settings from the ODES main
menu.
Please note that NO testing can be performed unless the equipment is calibrated and
verified. This is especially true if you select the Must be done option in Administration
ŇEnvironment Setting. No tests can be performed until the calibration is complete.