User's Manual

58
You are now ready to attach the calibration tool in order to finish manually calibrating the Algometer:
Calibration Tool
The calibration tool (above) has a weight that needs to be taken into account. The Algometer calibration
tool weighs 0.3 pounds. Add this value to the total amount being applied to the tool. For example, the
Algometer requires that a minimum of 25 pounds be used. If you choose to calibrate with this weight plus
the calibration tool, the total amount of weight to enter in the blank field would be 25.3 lbs.
Place the weight being used on top of the attachment.
Once the weight is steady, click OK to complete calibration.