Universal Task Master
When I go into a UTM test, I get a message that says “Click OK to Calibrate” but I just calibrated
my UTM.
ODES and the UTM perform a self-calibration each time you use the UTM load cells. Since you calibrate
your UTM system without any accessories attached, the computer needs to first measure the accessory
prior to commencing a test. This will re-zero the load cell to include the accessory. Once you click OK,
you may proceed with the test.
One of the UTM load cell readings is not correct.
Ensure the cable connections are tight at the UTM center.
Ensure the cable connections are tight at the Data Acquisition Box.
Re-calibrate the load cell.
Check the Data Acquisition Box troubleshooting section.
The UTM handle rotation readings are not correct.
Ensure the cable connection is tight at the UTM center.
Ensure the cable connections are tight at the Data Acquisition Box.
Re-calibrate the handle rotation.
Check the Data Acquisition Box troubleshooting section.
The UTM force scale distance readings are not correct.
Ensure the cable connection is tight at the UTM center.
Ensure the cable connections are tight at the Data Acquisition Box.
Re-calibrate the force scale distances.
Check the Data Acquisition Box troubleshooting section.
Cannot select Start Test for one of the UTM tests.
Ensure the cable connection is tight at the UTM center.
Ensure the cable connections are tight at the Data Acquisition Box.
Re-calibrate the UTM.
Check the Data Acquisition Box troubleshooting section.
Exit the screen and re-enter the test.
Cannot see the Universal Task Master calibration icon in the Utilities menu.
The Universal Task Master option is not installed on your computer. Call the Customer Service
Department at BTE Medical.
Cannot select the Universal Task Master icon when I try to create a new strength test or work
simulation test.
The Universal Task Master option is not installed on your computer. Call the technical support
department at BTE Medical.