User's Manual
201   
Palmar Pinch: 
The Palmar Pinch Grip strength test is performed per the protocol outlined in published research (see 
ODES CD for Pinch Strength Protocol) with a Pinch Dynamometer. This Pinch Grip dynamometer 
measures strength in the Palmar Pinch position of both the right and left hands. 
Click Palmar to perform a Palmar Pinch Grip Test. 
The “Pinch Grip – Palmar” screen has two grids for real-time graphs of the left and right hand strength 
curves. Each grid contains three reference lines that represent the normal values as well as upper and 
lower limits for the client, based on their age, gender and hand dominance. This information is obtained 
through the information the evaluator provides in the Client Information screen of ODES. The correct 
gender, date of birth and hand dominance of the client must
 be entered when performing these tests to 
ensure accurate and complete data analysis. 
Instruct the client to hold the pinch dynamometer between the tip of the thumb and the tips of the index 
and middle fingers. 
The client should be in a seated position as follows: 
Both feet flat on the floor 
Upper arm next to body 
Elbow flexed at 90 degrees 
Hand & forearm in slight shoulder internal rotation (toward the centre front of the torso) 
Forearm should not be resting on any surface while pinching 
Click Start Test to start the Palmar Pinch Grip Test. 