User's Manual
158
Elbow Flexion 
With the client in a standing position, manually landmark the lateral forearm with the hand in a supinated 
position. Using only the Master sensor, align it in the sagittal plane. Click the button on the Master sensor 
to take the initial reading. Have the client maximally flex the elbow. Click the button on the Master sensor 
a second time to take the final reading. ODES will automatically document the final true range. Return the 
client's elbow to the neutral position. Repeat these steps two more times on the same side and then 
repeat 3 times for the other side. 
Wrist Pronation 
With the client in a standing position, manually landmark the distal radial head with the wrist in a neutral 
position. Using only the Master sensor, align it in the coronal plane. Click the button on the Master sensor 
to take the initial reading. Have the client maximally pronate the wrist. Click the button on the Master 
sensor a second time to take the final reading. ODES will automatically document the final true range. 
Return the client's wrist to the neutral position. Repeat these steps two more times on the same side and 
then repeat 3 times for the other side. 