User's Manual Part 1

page
4
section 05
40040005 rev. 000
Select the Cervical icon from the main calibration screen to bring up the MCU™ calibration
screen (Figure 5-3).
The screen will always indicate the last day of successful calibration in addition to the amount
of weight used for the last calibration.
Step 1. Secure the head brace you wish to calibrate to the calibration plate. Make sure the
RJ45 cable is plugged into the head brace as well to the jack closest to the front of the unit
on the top of the MCU™ (Figure 5-4). Click the first OK on the calibration screen.
Step 2. Place the calibration block on the head brace (Figure 5-5).
Figure 5-3. Cervical Calibration Screen
Figure 5-4. Preparing Head Brace for Calibration
Figure 5-5. Placing Calibration Block on Head Brace
0.7lb
0.3kg