3.19.11 Cluster Prediction Protocol
Sets/changes the filter for the Prediction Protocol of Cluster and SIMCA calibrations.
Filter Settings
not in a cluster (!)
residual is within limits, but the distance is higher than allowed (spectrum is
not in a cluster), default: on.
false identified (?)
residual is within limits, distance is smaller than allowed (spectrum is in a
cluster), the predicted and original property name are different: this spectrum
is identified as false, default: on.
identified (*)
residual is within limits, distance is smaller than allowed, the predicted and
original property names are the same: this spectrum is identified correctly,
default: off.
not identified CLU
OK (%)
residual is not within limits, distance is smaller than allowed, the predicted
and original property names do not match: this known spectrum is not
identified, default: on.
not identified CLU
BAD (&)
residual is not within limits, distance is smaller than allowed, the predicted
and original property names are not the same: this spectrum is not identified,
but it is in a known cluster, default: on.
not identified
known (-)
residual is not within limits, the distance is not fulfilled, but the property is
known in the calibration, default: on.
not identified
unknown (=)
residual and distance is not fulfilled, the property is not available in the
calibration, default: on.
With the shown settings, only correct identified spectra are not listed in the prediction protocol.